Abstract
Background: Balloon atrial septostomy (BAS) is an emergent and essential cardiac intervention to enhance intercirculatory mixing at atrial level in deoxygenated patients diagnosed with transposition of the great arteries (TGA) and restrictive foramen ovale. The recent recall of several BAS catheters and the changes in the European legal framework for medical devices (MDR 2017/745), has led to an overall scarcity of BAS catheters and raised questions about the use, safety, and experience of the remaining NuMED Z-5 BAS catheter.
Aims: To evaluate and describe the practice and safety of the Z-5 BAS catheter, and to compare it to the performance of other BAS catheters.
Methods: A retrospective single-center cohort encompassing all BAS procedures performed with the Z-5 BAS catheter in TGA patients between 1999 and 2022.
Results: A total of 182 BAS procedures were performed in 179 TGA-newborns at Day 1 (IQR 0-5) days after birth, with median weight of 3.4 (IQR 1.2-5.7) kg. The need for BAS was urgent in 90% of patients. The percentage of BAS procedures performed at bedside increased over time from 9.8% (before 2010) to 67% (2017-2022). Major complication rate was 2.2%, consisting of cerebral infarction (1.6%) and hypovolemic shock (0.5%). The rate of minor complications was 9.3%, including temporary periprocedural AV-block (3.8%), femoral vein thrombosis (2.7%), transient intracardiac thrombus (0.5%), and atrial flutter (2.2%). BAS procedures performed at bedside and in the cardiac catheterization laboratory had similar complication rates.
Conclusions: BAS using the Z-5 BAS catheter is both feasible and safe at bedside and at the cardiac catheterization laboratory with minimal major complications.
Keywords: balloon atrial septostomy; cardiac catheterization; congenital heart disease; septostomy catheter; transposition of the great arteries.
© 2023 The Authors. Catheterization and Cardiovascular Interventions published by Wiley Periodicals LLC.
Comment in
-
Balloons, sheaths, and complications in balloon atrial septostomy.Catheter Cardiovasc Interv. 2024 Jun;103(7):1171-1172. doi: 10.1002/ccd.31025. Epub 2024 Mar 28. Catheter Cardiovasc Interv. 2024. PMID: 38545703 No abstract available.
-
Reply to Letter to the Editor: "Balloons, sheaths, and complications in balloon atrial septostomy".Catheter Cardiovasc Interv. 2024 Jul;104(1):82-83. doi: 10.1002/ccd.31072. Epub 2024 May 9. Catheter Cardiovasc Interv. 2024. PMID: 38720664 No abstract available.
References
REFERENCES
-
- Boehm W, Emmel M, Sreeram N. Balloon atrial septostomy: history and technique. Images Paediatr Cardiol. 2006;8(1):8-14.
-
- Cinteza E, Carminati M. Balloon atrial septostomy-almost half a century after. Maedica. 2013;8(3):280-284.
-
- Szymanski MW, Moore SM, Kritzmire SM, Goyal A. Transposition of the great arteries. 2023 Jan 15. StatPearls [Internet]. StatPearls Publishing; 2023.
-
- Rashkind WJ. Creation of an atrial septal defect without thoracotomy. A palliative approach to complete transposition of the great arteries. JAMA. 1966;196(11):991-992.
-
- Matter M, Almarsafawy H, Hafez M, Attia G, Elkhier MMA. Balloon atrial septostomy: the oldest cardiac interventional procedure in Mansoura. Egypt Heart J. 2011;63(2):125-129.