The sensitivity of the stiffness and thickness of a titanium inlay in a cementless PEEK femoral component to the micromotions and bone strain energy density

Affiliations
Free article

The sensitivity of the stiffness and thickness of a titanium inlay in a cementless PEEK femoral component to the micromotions and bone strain energy density

Corine E Post et al. Med Eng Phys. 2023 Dec.
Free article

Abstract

Polyetheretherketone (PEEK) has been proposed as alternative material for total knee arthroplasty implants due to its low stiffness, which may reduce stress-shielding. In cementless fixation, a proper primary fixation is required for long-term fixation. Previous research showed that the lower stiffness of a cementless PEEK femoral component results in larger micromotions at the implant-bone interface compared to a cobalt-chrome femoral component. A titanium inlay on the PEEK implant surface may improve the primary fixation while maintaining the favourable stiffness properties. Therefore, the effect of thickness and stiffness of a titanium inlay on the primary fixation and stress-shielding was investigated. A finite element model of the femur and femoral component was created with five titanium inlay variants. The micromotions and strain energy density (SED) were quantified as outcome measures. The distal thin - proximal thick variant showed the largest resulting micromotions (51.2 µm). Relative to the all-PEEK femoral component, the addition of a titanium inlay reduced the micromotions with 30 % to 40 % without considerably affecting the stress-shielding capacity (strain energy difference of 6 % to 10 %). Differences in micromotions (43.0-51.2 µm) and SED between the variants were relatively small. In conclusion, the addition of a titanium inlay could lead to a reduction of the micromotions without substantially affecting the SED distribution.

Keywords: Cementless femoral component; Finite element analysis; Micromotions; Polyetheretherketone; Strain energy density; Titanium inlay.

PubMed Disclaimer

Conflict of interest statement

Declaration of Competing Interest The authors declare the following financial interests/ personal relationships which may be considered as potential competing interests: A.B. is a paid employee of Invibio Ltd. and had a role in the study design and review of the manuscript. N.V. is a paid consultant for Invibio Ltd. D.J. is a paid consultant for Invibio Ltd. and receives research support as a principal investigator from Invibio Ltd. The other authors declare no conflict of interest.