m-PIMA™ HIV1/2 VL: A suitable tool for HIV-1 and HIV-2 viral load quantification in West Africa
- PMID: 38128833
- DOI: 10.1016/j.jviromet.2023.114872
m-PIMA™ HIV1/2 VL: A suitable tool for HIV-1 and HIV-2 viral load quantification in West Africa
Abstract
Point-of-Care for HIV viral RNA quantification seems to be a complementary strategy to the existing conventional systems. This study evaluated the performance of the m-PIMA™ HIV1/2 Viral Load for the quantification of both HIV-1 and HIV-2 RNA viral load. A total of 555 HIV-1 and 90 HIV-2 samples previously tested by Abbott RealTime HIV-1 (Abbott, Chicago, USA) and Generic HIV-2® Charge virale (Biocentric, France) were tested using the m-PIMA™ HIV1/2 Viral Load at the HIV National Reference lab in Senegal. For HIV-1, Pearson correlation and Bland-Altman plots showed a coefficient r = 0.97 and a bias of -0.11 log10 copies/ml (95% confidence interval [CI]: -0.086 to -0.133 log10 copies/ml) for the m-PIMA™ HIV1/2 Viral Load, respectively. Sensitivity and specificity at 3 log10 copies/ml (threshold of virological failure) were 93.6% (95%[CI]: 91.5% to 95.6%) and 99.1% (95%[CI]: 98.3% to 99.9%), respectively. For HIV-2, a correlation of r = 0.95 was also noted with a bias of - 0.229 log10 copies/ml (95%[CI]: -0.161 to -0.297 log10 copies/ml). Sensitivity and specificity at 3 log10 copies/ml were 97.6% (95%[CI]: 94.3% to 100%) and 93.9% (95%[CI]: 88.9% to 98.8%), respectively. These results confirmed that m-PIMA™ HIV1/2 VL could be a good alternative for HIV-1 and HIV-2 viral load testing in decentralized settings in Senegal.
Keywords: HIV-1; HIV-2; POC; Plasma Viral load; Senegal.
Copyright © 2023 The Authors. Published by Elsevier B.V. All rights reserved.
Conflict of interest statement
Declaration of Competing Interest The authors do not declare any conflict of interest.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous