Single-Use Lens Compromise During Selective Laser Trabeculoplasty
- PMID: 38129939
- DOI: 10.1097/IJG.0000000000002351
Single-Use Lens Compromise During Selective Laser Trabeculoplasty
Abstract
Prcis: Damage to disposable selective laser trabeculoplasty (SLT) lenses correlated with the amount of total laser energy. Furthermore, greater lens damage was associated with diminished patient treatment response.
Purpose: Disposable SLT lenses have been observed to be damaged during use by laser, potentially affecting therapeutic response. This study sought to identify factors associated with the magnitude of lens damage and its effect on treatment outcomes.
Patient and methods: We analyzed 113 eyes from 82 patients who underwent SLT between 2020 and 2021 at an American academic medical center. For each procedure, we recorded baseline patient characteristics, procedure settings, physician age (as a possible factor causing laser defocus due to accommodation), and area of lens damage. Treatment response was calculated as the difference between preoperative and postoperative intraocular pressure (IOP).
Results: Area of lens damage was associated with greater total laser energy ( r =0.34, P <0.001) and greater mean energy per application ( r =0.37, P <0.001). IOP reduction decreased with increasing lens damage ( r =-0.19, P =0.022). Lens damage was not associated with patient characteristics indicative of glaucoma severity or user settings (physician age and slit lamp used). Greater IOP reduction was associated with higher preop IOP ( r =0.46, P <0.001) and a smaller cup-to-disc ratio ( r =0.22, P =0.036). IOP reduction was not associated with retinal nerve fiber layer thickness, patient age, or the number of glaucoma medications.
Conclusions: Increased laser damage to disposable SLT lenses was associated with diminished treatment effect. The amount of damage was correlated with total and average laser energy, but not patient or physician factors. Overall, these findings call for reconsideration of using disposable SLT lenses as part of routine practice.
Copyright © 2023 Wolters Kluwer Health, Inc. All rights reserved.
Conflict of interest statement
Disclosure: The authors declare no conflict of interest.
References
-
- Gazzard G, Konstantakopoulou E, Garway-Heath D, et al. Selective laser trabeculoplasty versus eye drops for first-line treatment of ocular hypertension and glaucoma (LiGHT): a multicentre randomised controlled trial. Lancet. 2019;393:1505–1516.
-
- Gazzard G, Konstantakopoulou E, Garway-Heath D, et al. Laser in Glaucoma and Ocular Hypertension (LiGHT) Trial: six-year results of primary selective laser trabeculoplasty versus eye drops for the treatment of glaucoma and ocular hypertension. Ophthalmology. 2023;130:139–151.
-
- Junk AK, Chen PP, Lin SC, et al. Disinfection of tonometers: a report by the American Academy of Ophthalmology. Ophthalmology. 2017;124:1867–1875.
-
- Lee B, Szirth BC, Fechtner RD, et al. Are disposable and standard gonioscopy lenses comparable? J Glaucoma. 2017;26:e157–e159.
-
- Mukhtar S, Shazly T. Laser-induced damage to disposable single-mirror trabeculoplasty lens during selective laser trabeculoplasty. J Glaucoma. 2020;29:33–34.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous