Influence of Embedding Medium on iPex Electronic Apex Locator Accuracy: An Ex Vivo Comparative Study Using Alginate, Saline, and Gelatin Models
- PMID: 38152904
- DOI: 10.5005/jp-journals-10024-3583
Influence of Embedding Medium on iPex Electronic Apex Locator Accuracy: An Ex Vivo Comparative Study Using Alginate, Saline, and Gelatin Models
Abstract
Aim: Accurate working length determination during root canal treatment is essential for achieving successful outcomes. This study aimed to evaluate the impact of embedding medium on the accuracy of iPex electronic apex locator (EAL).
Materials and methods: Sixty-one extracted single-rooted teeth were decoronated and coronally flared with Gates-Glidden burs. Actual canal length (ACL) was obtained by introducing a size 8 K-file until its tip reached the most coronal border of the apical foramen. This step was performed thrice and then averaged. Deducting 0.5 mm from the ACL provided the working length (WL). The teeth were randomly placed in plastic containers filled with freshly mixed alginate, gelatin, or saline, with the lip clip placed in the medium. The blinded operator obtained electronic measurements using iPex by advancing a K-file with a size compatible with the canal attached to the file clip and advanced until the 0.0 mark, then withdrawn to the 0.5 mark. This step was performed thrice and then averaged. Data were analyzed using ANOVA and Tukey's post hoc test, with significance level set at 5% (α = 0.05).
Results: The mean difference between WL and iPex length obtained in the gelatin model was significantly longer than the difference with mean iPex length in alginate (p = 0.005) and in saline (p < 0.001). There was no significant difference between iPex readings obtained in alginate and saline (p = 0.249).
Conclusion: The use of freshly mixed alginate or saline for ex vivo assessment of iPex is recommended, whereas the use of gelatin could increase the chances of readings longer than looked for.
Clinical significance: Identifying the optimum embedding medium for ex vivo testing of EALs permits the comparison and assessment of several factors affecting EALs' precision under standardized conditions. This helps in understanding EAL performance in vivo and in optimizing its clinical utilization.
Keywords: Apex locators; Electronic apex locator; Electronic working length; Embedding medium; Endodontics; iPex Working length determination..
Similar articles
-
A laboratory study of the accuracy of three electronic apex locators: influence of embedding media and radiographic assessment of the electronic apical limit.Int Endod J. 2021 Jul;54(7):1200-1206. doi: 10.1111/iej.13507. Epub 2021 Mar 30. Int Endod J. 2021. PMID: 33683733
-
Influence of Hand File Size on the Accuracy of Root ZX and iPex Electronic Apex Locators: An In Vitro Study.Cureus. 2023 May 29;15(5):e39662. doi: 10.7759/cureus.39662. eCollection 2023 May. Cureus. 2023. PMID: 37256168 Free PMC article.
-
In vivo evaluation of the iPex and Root ZX electronic apex locators using various irrigants.Int Endod J. 2013 Aug;46(8):769-74. doi: 10.1111/iej.12057. Epub 2013 Mar 28. Int Endod J. 2013. PMID: 23551276 Clinical Trial.
-
Electronic apex locator.Dent Clin North Am. 2004 Jan;48(1):35-54. doi: 10.1016/j.cden.2003.10.005. Dent Clin North Am. 2004. PMID: 15066506 Review.
-
Efficacy of electronic apex locators in comparison with intraoral radiographs in working length determination- a systematic review and meta-analysis.BMC Oral Health. 2024 May 4;24(1):532. doi: 10.1186/s12903-024-04259-w. BMC Oral Health. 2024. PMID: 38704529 Free PMC article.