Multiscale Monte Carlo simulations for dosimetry in x-ray breast imaging: Part I - Macroscopic scales
- PMID: 38156766
- DOI: 10.1002/mp.16910
Multiscale Monte Carlo simulations for dosimetry in x-ray breast imaging: Part I - Macroscopic scales
Abstract
Background: X-ray breast imaging modalities are commonly employed for breast cancer detection, from screening programs to diagnosis. Thus, dosimetry studies are important for quality control and risk estimation since ionizing radiation is used.
Purpose: To perform multiscale dosimetry assessments for different breast imaging modalities and for a variety of breast sizes and compositions. The first part of our study is focused on macroscopic scales (down to millimeters).
Methods: Nine anthropomorphic breast phantoms with a voxel resolution of 0.5 mm were computationally generated using the BreastPhantom software, representing three breast sizes with three distinct values of volume glandular fraction (VGF) for each size. Four breast imaging modalities were studied: digital mammography (DM), contrast-enhanced digital mammography (CEDM), digital breast tomosynthesis (DBT) and dedicated breast computed tomography (BCT). Additionally, the impact of tissue elemental compositions from two databases were compared. Monte Carlo (MC) simulations were performed with the MC-GPU code to obtain the 3D glandular dose distribution (GDD) for each case considered with the mean glandular dose (MGD) fixed at 4 mGy (to facilitate comparisons).
Results: The GDD within the breast is more uniform for CEDM and BCT compared to DM and DBT. For large breasts and high VGF, the ratio between the minimum/maximum glandular dose to MGD is 0.12/4.02 for DM and 0.46/1.77 for BCT; the corresponding results for a small breast and low VGF are 0.35/1.98 (DM) and 0.63/1.42 (BCT). The elemental compositions of skin, adipose and glandular tissue have a considerable impact on the MGD, with variations up to 30% compared to the baseline. The inclusion of tissues other than glandular and adipose within the breast has a minor impact on MGD, with differences below 2%. Variations in the final compressed breast thickness alter the shape of the GDD, with a higher compression resulting in a more uniform GDD.
Conclusions: For a constant MGD, the GDD varies with imaging modality and breast compression. Elemental tissue compositions are an important factor for obtaining MGD values, being a source of systematic uncertainties in MC simulations and, consequently, in breast dosimetry.
Keywords: Monte Carlo; breast CT; contrast enhanced digital mammography; digital breast tomosynthesis; digital mammography; radiation dosimetry.
© 2023 The Authors. Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine.
References
REFERENCES
-
- Gold RH, Bassett LW, Widoff BE. Highlights from the history of mammography. RadioGraphics. 1990;10:1111-1131.
-
- European Society of Radiology. Screening & Beyond: Medical Imaging in the Detection, Diagnoses and Management of Breast Diseases. ESR; 2016.
-
- Marmot MG, Altman DG, Cameron DA, Dewar JA, Thompson SG, Wilcox M; The Independent UK Panel on Breast Cancer Screening. The benefits and harms of breast cancer screening: an independent review. Br J Cancer. 2013;108:2205-2240.
-
- Sardanelli F, Aase HS, Álvarez M, et al. Position paper on screening for breast cancer by the European Society of Breast Imaging (EUSOBI) and 30 national breast radiology bodies from Austria, Belgium, Bosnia and Herzegovina, Bulgaria, Croatia, Czech Republic, Denmark, Estonia, Finland, France, Germany, Greece, Hungary, Iceland, Ireland, Italy, Israel, Lithuania, Moldova, The Netherlands, Norway, Poland, Portugal, Romania, Serbia, Slovakia, Spain, Sweden, Switzerland and Turkey. Eur Radiol. 2017;27:2737-2743.
-
- Skaane P, Bandos AI, Gullien R, et al. Comparison of digital mammography alone and digital mammography plus tomosynthesis in a population-based screening program. Radiology. 2013;267:47-56.
MeSH terms
Grants and funding
- 2015/21873-8/Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)
- 2018/05982-0/Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)
- 2021/08923-7/Fundação de Amparo à Pesquisa do Estado de São Paulo (FAPESP)
- 140155/2019-8/Conselho Nacional de Desenvolvimento Científico e Tecnológico
- 313600/2021-0/Conselho Nacional de Desenvolvimento Científico e Tecnológico
- Finance Code 001/Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES)
- 2581/23/Fundo de apoio ao ensino, pesquisa e Extensão da UNICAMP
- Emerging leaders of the Americas program of the Government of Canada
- 06267-2016/Natural Sciences and Engineering Research Council of Canada
- Canada Research Charis
LinkOut - more resources
Full Text Sources
Medical