Mitigating slice cross-talk in multi-slice multi-echo spin echo T2 mapping
- PMID: 38156822
- DOI: 10.1002/mrm.29987
Mitigating slice cross-talk in multi-slice multi-echo spin echo T2 mapping
Abstract
Purpose: To investigate whether a T2 inter-slice variation could occur when a multi-slice multi-echo spin echo (MESE) sequence is used for image acquisition and to propose an enhanced method for reconstructing T2 maps that can effectively address and correct these variations.
Methods: Bloch simulations were performed accounting for the direct saturation effect to evaluate magnetization changes in multi-slice 2D MESE sequence. Experimental phantom scans were performed to validate these simulations. An improved version of the dictionary-based reconstruction approach was proposed, enabling the creation of a multi-slice dictionary of echo modulation curves (EMC). The corresponding method has been assayed considering inter-slice T2 variation with phantoms and in lower leg.
Results: Experimental and numerical study illustrate that direct saturation leads to a bias of EMCs. This bias during the T2 maps reconstructions using original single-slice EMC-dictionary method led to inter-slice T2 variation of 2.03% in average coefficient of variation (CV) in agarose phantoms, and up to 2.8% in vivo (for TR = 2 s, slice gap = 0%). A reduction of CV was observed when increasing the gap up to 100% (0.36% in phantoms, and up to 1.5% in vivo) or increasing TR up to 4 s (0.76% in phantoms, and up to 1.9% in vivo). Matching the multi-slice experimental data with multi-slice dictionaries provided a reduced CV of 0.54% in phantoms and up to 2.3% in vivo.
Conclusion: T2 values quantified from multi-slice MESE images using single-slice dictionaries are biased. A dedicated multi-slice EMC method providing the appropriate dictionaries can reduce the inter-slice T2 variation.
Keywords: T2 mapping; direct saturation effect; echo modulation curve; inter-slice T2 variation; quantitative MRI; slice cross-talk.
© 2023 International Society for Magnetic Resonance in Medicine.
References
REFERENCES
-
- Tahir E, Sinn M, Bohnen S, et al. Acute versus chronic myocardial infarction: diagnostic accuracy of quantitative native T1 and T2 mapping versus assessment of edema on standard T2-weighted cardiovascular MR images for differentiation. Radiology. 2017;285:83-91.
-
- Kim HK, Laor T, Horn PS, Racadio JM, Wong B, Dardzinski BJ. T2 mapping in Duchenne muscular dystrophy: distribution of disease activity and correlation with clinical assessments. Radiology. 2010;255:899-908.
-
- David-Vaudey E, Ghosh S, Ries M, Majumdar S. T2 relaxation time measurements in osteoarthritis. Magn Reson Imaging. 2004;22:673-682.
-
- Majumdar S, Orphanoudakis SC, Gmitro A, O'Donnell M, Gore JC. Errors in the measurements of T2 using multiple-echo MRI techniques. I. Effects of radiofrequency pulse imperfections. Magn Reson Med. 1986;3:397-417.
-
- Radunsky D, Blumenfeld-Katzir T, Volovyk O, et al. Analysis of magnetization transfer (MT) influence on quantitative mapping of T2 relaxation time. Magn Reson Med. 2019;82:145-158.
MeSH terms
Grants and funding
- Excellence Initiative of Aix-Marseille University-A*MIDEX, a French "Investissements d'Avenir" Program through the Multiwave Chair of Medical Imaging
- French Embassy in Russia (Metchnikov Program)
- HШ2359.2022.4/The Ministry of Science and Higher Education of the Russian Federation, grant for Scientific School No
- 075-15-2021-592/The Ministry of Science and Higher Education of the Russian Federation Project No
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous