. 2023 Dec 1;94(12):125113.
doi: 10.1063/5.0176783.

Laminar-type gratings overcoated with carbon-based materials to enhance analytical sensitivity of flat-field emission spectrograph in the VUV region

Affiliations

Laminar-type gratings overcoated with carbon-based materials to enhance analytical sensitivity of flat-field emission spectrograph in the VUV region

T Murano et al. Rev Sci Instrum. .

Abstract

Laminar-type spherical diffraction gratings overcoated with carbon-based materials were designed, fabricated, and evaluated for the purpose of enhancing the analytical sensitivity of the flat-field spectrograph in a vacuum ultraviolet region of 35-110 eV. As the design benchmark for numerical calculations, diffraction efficiency (DE) and spectral flux, which are defined by the product of the DE and numerical aperture and correlate with the analytical sensitivity of the spectrograph, were used. To simplify the feasibility study on the overcoating effects, we assumed a laminar-type grating having a grating constant of 1/1000 mm and coated with a Au layer of 30.0 nm thickness and an incidence angle of 84.0°. The optimized groove depth and duty ratio were 30.0 nm and 0.3, respectively. In addition, the optimum thicknesses of the overcoating layer were 44, 46, 24, and 30 nm for B4C, C, diamond-like-carbon, and SiC, respectively. Based on these results, we have fabricated a varied-line-spacing holographic grating overcoated with B4C with a thickness of 47 nm. For the experimental evaluation, we used the light source of Mg-L and Al-L emissions excited by the electron beam generated from an electron microscope, an objective flat-field spectrograph, and a CCD imaging detector. The experimental results showed that the spectrograph employing a new grating overcoated with the B4C layer indicated almost the same spectral resolution and 2.9-4.2 times higher analytical sensitivity compared with those obtained with a previously designed Au-coated grating having a grating constant of 1/1200 mm and used at an incidence of 86.0°.

PubMed Disclaimer