Effects of Apex Size on Dissolution Profiles in the USP II Paddle Apparatus

Affiliations

Effects of Apex Size on Dissolution Profiles in the USP II Paddle Apparatus

Hiroyuki Yoshida et al. AAPS PharmSciTech. .

Erratum in

Abstract

The use of apex vessels may solve coning problems associated with dissolution testing. However, excessive dissolution acceleration can reduce the discriminatory power. This study aimed to clarify how different apex vessel sizes affect the dissolution behavior of cone-forming formulations. Five apex vessels with different heights, centralities, and compendial vessels were used. The paddle rotation speed at which the coning phenomenon resolved was measured using standard particles of different densities. Three model formulations-USP prednisone tablets, atorvastatin calcium hydrate tablets, and levofloxacin fine granules-were selected, and dissolution tests were conducted at 30-100 revolutions per minute (rpm). Compared to the compendial vessels, the disappearance of standard particles at the apex base at lower paddle speeds in apex vessels was observed. Standard particles tended to remain in the center of the apex vessels and disappear at rotational speeds comparable to those of the compendial vessels. Dissolution increased in an apex height-dependent manner in the model formulations, except for the atorvastatin calcium hydrate tablets at 50 rpm. For levofloxacin fine granules, dissolution was also improved by reducing the paddle agitation speed to 30 rpm in the compendial vessels. Differences in apex centrality by 3 mm did not affect the dissolution rate. Our results indicate that apex vessels with low apex heights have a mount-resolving effect, but the degree of dissolution improvement by avoiding the coning phenomenon depends on the formulation characteristics used in the dissolution tests.

Keywords: apex size; apex vessel; coning; dissolution testing; paddle apparatus.

PubMed Disclaimer

References

    1. International Conference on Harmonisation; Guidance on Q4B evaluation and recommendation of pharmacopoeial texts for use in the International Conference on Harmonisation Regions; Annex 7 on Dissolution Test General Chapter; availability. Notice. Fed Regist. 2010;75:17148–9.
    1. Suarez-Sharp S, Delvadia PR, Dorantes A, Duan J, Externbrink A, Gao Z, et al. Regulatory perspectives on strength-dependent dissolution profiles and biowaiver approaches for immediate release (IR) oral tablets in new drug applications. AAPS J. 2016;18:578–88. https://doi.org/10.1208/s12248-016-9893-2 . - DOI - PubMed - PMC
    1. Langer C, Diaz DA, Van Alstine L, Bandi N, Fitzgerald S, Colgan S. Analysis of the regulatory divergence of comparative dissolution requirements for post-approval manufacturing changes of modified release solid oral dosage forms. AAPS J. 2019;22:2. https://doi.org/10.1208/s12248-019-0383-1 . - DOI - PubMed
    1. Abend AM, Zhang L, Fredro-Kumbaradzi E, Hoffelder T, Cohen MJ, Anand O, et al. Current approaches for dissolution similarity assessment, requirements, and global expectations. AAPS J. 2022;24:50. https://doi.org/10.1208/s12248-022-00691-4 . - DOI - PubMed
    1. International Council for Harmonisation of Technical Requirements for Pharmaceuticals for Human Use (ICH) M9: Biopharmaceutics classification system-based biowaiver. Available from: https://www.ich.org/page/multidisciplinary-guidelines . Accessed 7 Sept 2023.

LinkOut - more resources