Performance evaluation of the Sysmex XN-1000V in side-by-side comparison with the Siemens ADVIA 120 and manual methods for healthy CD Sprague-Dawley rats and CD-1 mice
- PMID: 38164989
- DOI: 10.1111/vcp.13314
Performance evaluation of the Sysmex XN-1000V in side-by-side comparison with the Siemens ADVIA 120 and manual methods for healthy CD Sprague-Dawley rats and CD-1 mice
Abstract
Background: The Sysmex XN-1000V automated hematology analyzer with multispecies software was released in June 2017 for use in research laboratories. Laser light, impedance, fluorescent staining, and fluorescent flow cytometry are used to analyze whole blood for CBC, reticulocyte counts, and WBC counts, including a 5-part differential leukocyte analysis.
Objectives: A side-by-side comparison of the Sysmex XN-1000V with the Siemens ADVIA 120 in analyzing blood from healthy mice and rats will provide insight into the performance of the new analyzer and its capabilities for use in drug development studies. Method correlation analyses on normal mouse and rat hematology data collected with both analyzers and manual reference methods will help determine the reliability of the data produced using the Sysmex XN-1000V analyzer.
Methods: Whole blood samples collected in K2 EDTA from healthy CD-1 mice and CD Sprague-Dawley rats were analyzed in parallel with the XN-1000V and ADVIA 120 analyzers. Male and female mice, approximately 6-9 weeks old, and male and female rats, approximately 7-9 weeks old, were included in this study. Manual reference methods for WBC differential leukocyte analysis and packed cell volume (PCV) measurements were also performed. EP Evaluator version 11.2 (Data Innovations LLC, South Burlington, VT, USA) was used for method comparison statistical analysis.
Results: Most hematologic parameters for naïve mice and rats achieved correlation in the fair to excellent range, with the majority showing very good to excellent correlation with low biases (<11.0%) for cohorts analyzed separately and when cohort data were combined.
Conclusions: The Sysmex XN-1000V Hematology Analyzer provided comparable results to those obtained from the Siemens ADVIA 120. We found the Sysmex XN-1000V Hematology Analyzer to be acceptable for use in drug development studies for rats and mice.
Keywords: automated analyzer; correlation; hematology; instrument evaluation; precision; stability.
© 2024 American Society for Veterinary Clinical Pathology.
Similar articles
-
Evaluation of blood erythroid parameters in male broiler chickens (Ross 308) with the Sysmex XT-2000iV and Sysmex XN-1000V analyzers and determination of hematological reference intervals obtained with manual and instrumental methods.Vet Clin Pathol. 2025 Jun;54(2):106-119. doi: 10.1111/vcp.70009. Epub 2025 Jun 3. Vet Clin Pathol. 2025. PMID: 40462459 Free PMC article.
-
Validation of the Sysmex XN-V hematology analyzer for canine specimens.Vet Clin Pathol. 2021 Jun;50(2):184-197. doi: 10.1111/vcp.12936. Epub 2021 Jun 21. Vet Clin Pathol. 2021. PMID: 34152026 Free PMC article.
-
Automated hematological cell count using sysmex XN-1000V in Testudo hermanni: Agreement with manual count.Res Vet Sci. 2024 Mar;169:105164. doi: 10.1016/j.rvsc.2024.105164. Epub 2024 Feb 1. Res Vet Sci. 2024. PMID: 38324973
-
Evaluation of canine and feline leukocyte differential counts obtained with the scil vCell 5 compared to the Advia 2120 hematology analyzer and a manual method.J Vet Diagn Invest. 2023 Nov;35(6):679-697. doi: 10.1177/10406387231187899. Epub 2023 Aug 23. J Vet Diagn Invest. 2023. PMID: 37612877 Free PMC article. Review.
-
Diagnostic role of Sysmex hematology analyzer in the detection of malaria: A systematic review and meta-analysis.PLoS One. 2024 Sep 6;19(9):e0296766. doi: 10.1371/journal.pone.0296766. eCollection 2024. PLoS One. 2024. PMID: 39240990 Free PMC article.
References
REFERENCES
-
- Schultze AE, Aulbach AD, Ennulat D. Clinical pathology in nonclinical toxicity testing. In: Haschek WM, Rousseaux CG, Wallig MA, Bolon B, eds. Haschek and Rousseaux's Handbook of Toxicologic Pathology. Vol 1. 4th ed. Academic Press (Elsevier); 2022:295-335.
-
- Aulbach AD, Ennulat D, Schultze AE. Interpretation of clinical pathology results in nonclinical toxicity testing. In: Haschek WM, Rousseaux CG, Wallig MA, Bolon B, eds. Haschek and Rousseaux's Handbook of Toxicologic Pathology. Vol 2. 4th ed. Academic Press (Elsevier); 2023:505-566.
-
- Aulbach AD, Ennulat D, Schultze AE. Clinical pathology in nonclinical toxicity studies. In: Faqi AS, ed. A Comprehensive Guide to Toxicology in Nonclinical Drug Development. 3rd ed. Academic Press (Elsevier); in press.
-
- Poitout-Belissent F, Aulbach A, Tripathi N, Ramaiah L. Reducing blood volume requirements for clinical pathology testing in toxicologic studies-points to consider. Vet Clin Pathol. 2016;45(4):534-551.
-
- Nabity MB, Harr KE, Camus MS, Flatland B, Vap LM. ASVCP guidelines: allowable total error hematology. Vet Clin Pathol. 2018;47(1):9-21.