Sensitivity and specificity of SS-OCT for detecting macular pathologies vs SD-OCT
- PMID: 38192061
- PMCID: PMC11045393
- DOI: 10.1097/j.jcrs.0000000000001394
Sensitivity and specificity of SS-OCT for detecting macular pathologies vs SD-OCT
Abstract
Purpose: To evaluate the sensitivity and specificity of swept-source optical coherence tomography (SS-OCT) biometer compared with the gold standard spectral-domain optical coherence tomography (SD-OCT) for detecting macular pathology in patients with cataract.
Setting: Eye Centers of Tennessee, Crossville, TN.
Design: Prospective, cross-sectional, observational, examiner-masked.
Methods: The study included 132 participants aged 50 years and older, who underwent precataract surgery work-up. All participants underwent fixation check retinal scans using SS-OCT biometer (IOLMaster 700) as well as full macular scans using Cirrus SD-OCT. 3 independent masked examiners evaluated the scans if they were normal or had a suspected pathology. Different measures of diagnostic accuracy were calculated for 3 examiners.
Results: True positive rate (sensitivity) ranged from 71.1% (32/45) to 79.2% (42/53), and false negative rate was between 20.8% (11/53) and 28.9% (13/45) for the 3 examiners. True negative rate (specificity) ranged from 86.8% (59/68) to 94.1% (64/68), and false positive rate was between 5.9 (4/68) and 13.2% (9/68). The fitted receiver operating characteristic area ranged from 0.83 to 0.95.
Conclusions: Using retinal SS-OCT biometer scans as a replacement of the dedicated macular SD-OCT for screening or diagnosing macular health would not be appropriate because of its low sensitivity. SS-OCT biometer may potentially fail to identify approximately one-fourth of patients who actually have the disease. Therefore, the final decision on macular health should be based on the gold standard SD-OCT scans. When full macular SD-OCT scans are not accessible, the limited retinal scan information from SS-OCT biometer may still provide useful insights into the macular health.
Copyright © 2024 The Author(s). Published by Wolters Kluwer Health, Inc. on behalf of ASCRS and ESCRS.
Figures
References
-
- Zafar S, Siddiqui MAR, Shahzad R, Shahzad MH. Swept-source optical coherence tomography to screen for macular pathology in eyes having routine cataract surgery. J Cataract Refract Surg 2017;43:324–327 - PubMed
-
- Guttman C. Pre-cataract surgery OCT means happier patient outcomes. Ophthalmology Times 2017. Available at: https://www.ophthalmologytimes.com/view/pre-cataract-surgery-oct-means-h.... Accessed July 14, 2020
-
- Klein BR, Brown EN, Casden RS. Preoperative macular spectral-domain optical coherence tomography in patients considering advanced-technology intraocular lenses for cataract surgery. J Cataract Refract Surg 2016;42:537–541 - PubMed
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical