Case Reports
. 2023 Oct 9;9(12):948-953.
doi: 10.1016/j.hrcr.2023.10.006. eCollection 2023 Dec.

Dispersion electrogram detection with an artificial intelligence software in redo paroxysmal atrial fibrillation ablation

Affiliations
Case Reports

Dispersion electrogram detection with an artificial intelligence software in redo paroxysmal atrial fibrillation ablation

Ngoda Manongi et al. HeartRhythm Case Rep. .
No abstract available

Keywords: Artificial intelligence; Atrial fibrillation; Catheter ablation; Driver; Pulmonary vein isolation; Spatiotemporal dispersion.

PubMed Disclaimer

Conflict of interest statement

Dr Goldbarg is on the Scientific Advisory Board of Volta Medical, and is the US principal investigator of the Tailored-AF Trial.

Figures

Figure 1
Figure 1
Spatiotemporal dispersion of multipolar electrograms (EGMs), detection with AI software (VX1; Volta Medical, Marseille, France), and catheter ablation. Left panel displays examples of dispersed EGMs from patient 1. EGMs are processed in real time by VX1 system (VX1 interface right panel) on 1.5-second signals and are considered dispersed when they display 1 continuously fractionated electrogram with a cumulative duration longer than 80% of the 1.5-second window. Stable intracardiac atrial dispersed electrograms are identified by VX1 algorithm on each dipole 1-2 to 19-20 (the dipoles schematized in the upper half cycle are blinking red). The lower part of the frame displays the average cycle length as recorded by the reference catheter (R) and the mapping catheter (M).
Figure 2
Figure 2
Electroanatomical maps from patients 1 (biatrial, A), 2 (left atrium, B), 3 (biatrial, C), and 4 (biatrial, D). Maps on the left demonstrate dispersion sites, where green points are highly dispersed and white points moderately dispersed. Right-sided maps show the dispersion and pulmonary vein isolation lesion sets. Blue dot in panels B and D shows point where termination of atrial fibrillation occurred.
Figure 3
Figure 3
Electroanatomical maps from patient 2. A: Image shows 43 left atrial points (white dots) mapped with the PENTARAY (Biosense Webster, Irvine, CA) and the VX1 (Volta Medical, Marseille, France) system. B: PENTARAY signals in the right carina, with dispersion demonstrated across the PENTARAY splines. PENTARAY 13/14 (white dot with highlighted edge/blue-green electrogram) is at the site of termination during ablation.

Similar articles

Cited by

References

    1. Chugh S.S., Havmoeller R., Narayanan K., et al. Worldwide epidemiology of atrial fibrillation: a Global Burden of Disease 2010 Study. Circulation. 2014;129:837–847. - PMC - PubMed
    1. Mody B.P., Raza A., Jacobson J., et al. Ablation of long-standing persistent atrial fibrillation. Ann Transl Med. 2017;5:305. - PMC - PubMed
    1. Verma A., Jiang C., Betts T.R., et al. Approaches to catheter ablation for persistent atrial fibrillation. N Engl J Med. 2015;372:1812–1822. - PubMed
    1. Verma A., Macle L. Persistent atrial fibrillation ablation: where do we go from here? Can J Cardiol. 2018;34:1471–1481. - PubMed
    1. Quintanilla J.G., Shpun S., Jalife J., Filgueiras-Rama D. Novel approaches to mechanism-based atrial fibrillation ablation. Cardiovasc Res. 2021;117:1662–1681. - PMC - PubMed

Publication types

LinkOut - more resources