Photon-Counting Detector CT: Clinical Utility of Virtual Monoenergetic Imaging Combined With Tin Prefiltration to Reduce Metal Artifacts in the Postoperative Ankle
- PMID: 38214560
- DOI: 10.1097/RLI.0000000000001058
Photon-Counting Detector CT: Clinical Utility of Virtual Monoenergetic Imaging Combined With Tin Prefiltration to Reduce Metal Artifacts in the Postoperative Ankle
Abstract
Objectives: The aim of this study was to compare the effectiveness and clinical utility of virtual monoenergetic image (VMI) reconstructions in computed tomography (CT) scans with and without tin prefiltration on a photon-counting detector (PCD) CT system to reduce metal implant artifacts in the postoperative ankle.
Materials and methods: This retrospective study included patients with internal fixation of the ankle scanned with and without tin prefiltration (Sn) on a PCD CT scanner between March and October 2023. Virtual monoenergetic images between 60 and 190 keV were reconstructed with a 10-keV increment in a bone kernel for both acquisitions (VMI Sn and VMI Std , respectively). Noise measurements assessed artifact reduction in the most prominent near-metal image distortions and were compared between acquisitions modes as well as between polychromatic images and VMIs. Three readers assessed the visibility of osseous healing along with interpretability and artifact extent for 5 reconstruction levels.
Results: A total of 48 patients (21 females, 27 males; mean age, 55.1 ± 19.4 years) were included in this study. Tin-prefiltered acquisitions (n = 30) had a lower artifact level for polychromatic images and VMIs compared with non-tin-prefiltered acquisitions (n = 18; P ≤ 0.043). A significant reduction of metal artifacts was observed for VMI Sn ≥120 keV compared with polychromatic images (hyperdense artifacts: 40.2 HU [interquartile range (IQR) 39.8] vs 14.0 HU [IQR 11.1]; P ≤ 0.01 and hypodense artifacts: 91.2 HU [IQR 82.4] vs 29.7 HU [IQR 39.6]; P ≤ 0.001). For VMI Std , this applied to reconstructions ≥100 keV (hyperdense artifacts: 57.7 HU [IQR 33.4] vs 19.4 HU [IQR 27.6]; P ≤ 0.001 and hypodense artifacts: 106.9 HU [IQR 76.1] vs 57.4 HU [IQR 55.7]; P ≤ 0.021). For visibility of osseous healing, VMI Sn at 120 keV yielded higher ratings compared with polychromatic images ( P ≤ 0.001), whereas image interpretability was rated better ( P = 0.023), and artifact extent was rated lower ( P ≤ 0.001) compared with polychromatic images.
Conclusions: Tin-prefiltered VMI at 120 keV showed a significant reduction in metal artifacts compared with polychromatic images, whereas visibility of osseous healing and image interpretability was improved. Therefore, tin-prefiltration PCD CT with VMI reconstructions may be a helpful complement to postsurgical CT imaging of the ankle in patients with metal implants.
Copyright © 2024 The Author(s). Published by Wolters Kluwer Health, Inc.
Conflict of interest statement
Conflicts of interest and sources of funding: The University Hospital Balgrist has an academic collaboration agreement with Siemens Healthineers, Bayer, and Balzano Informatik. Professor Reto Sutter receives royalties for his position as book editor at Breitenseher Publisher. The remaining authors do not have any conflict of interest to declare.
Similar articles
-
High-Pitch Multienergy Coronary CT Angiography in Dual-Source Photon-Counting Detector CT Scanner at Low Iodinated Contrast Dose.Invest Radiol. 2023 Sep 1;58(9):681-690. doi: 10.1097/RLI.0000000000000961. Invest Radiol. 2023. PMID: 36822655 Free PMC article.
-
Tungsten-Based Contrast Agent for Photon-Counting Detector CT Angiography in Calcified Coronaries: Comparison to Iodine in a Cardiovascular Phantom.Invest Radiol. 2024 Oct 1;59(10):677-683. doi: 10.1097/RLI.0000000000001073. Epub 2024 Mar 25. Invest Radiol. 2024. PMID: 38526041 Free PMC article.
-
Reduction of Streak Artifacts in the Superior Vena Cava for Better Visualization of Mediastinal Structures Through Virtual Monoenergetic Reconstructions Using a Photon-counting Detector Computed Tomography.J Thorac Imaging. 2025 Jul 1;40(4):e0822. doi: 10.1097/RTI.0000000000000822. J Thorac Imaging. 2025. PMID: 39885700
-
Rehabilitation for ankle fractures in adults.Cochrane Database Syst Rev. 2024 Sep 23;9(9):CD005595. doi: 10.1002/14651858.CD005595.pub4. Cochrane Database Syst Rev. 2024. PMID: 39312389
-
123I-MIBG scintigraphy and 18F-FDG-PET imaging for diagnosing neuroblastoma.Cochrane Database Syst Rev. 2015 Sep 29;2015(9):CD009263. doi: 10.1002/14651858.CD009263.pub2. Cochrane Database Syst Rev. 2015. PMID: 26417712 Free PMC article.
Cited by
-
Assessment of bone-implant interface image quality for in-vivo acetabular cup implants using photon-counting detector CT: Impact of tin pre-filtration.Eur J Radiol Open. 2025 Mar 27;14:100646. doi: 10.1016/j.ejro.2025.100646. eCollection 2025 Jun. Eur J Radiol Open. 2025. PMID: 40226312 Free PMC article.
-
The Clinical Value of the MAR + Metal Artifact Reduction Algorithm for Postoperative Assessment of Lumbar Internal Fixation.J Comput Assist Tomogr. 2025 Jul-Aug 01;49(4):675-681. doi: 10.1097/RCT.0000000000001724. Epub 2025 Jan 27. J Comput Assist Tomogr. 2025. PMID: 39876491 Free PMC article.
-
Assessment of metal artifacts from titanium wrist prostheses: photon-counting versus energy-integrating detector CT.Eur Radiol Exp. 2025 May 1;9(1):45. doi: 10.1186/s41747-025-00587-w. Eur Radiol Exp. 2025. PMID: 40310571 Free PMC article.
-
Postoperative Extremity Tomosynthesis-A Superimposition-Free Alternative to Standard Radiography?Invest Radiol. 2024 Nov 1;59(11):761-766. doi: 10.1097/RLI.0000000000001085. Epub 2024 May 7. Invest Radiol. 2024. PMID: 38709665 Free PMC article.
-
Photon-Counting Detector CT Applications in Musculoskeletal Radiology.Invest Radiol. 2025 Mar 1;60(3):198-204. doi: 10.1097/RLI.0000000000001108. Epub 2024 Aug 2. Invest Radiol. 2025. PMID: 39088264 Free PMC article. Review.
References
-
- Meyer E, Raupach R, Lell M, et al. Normalized metal artifact reduction (NMAR) in computed tomography. Med Phys . 2010;37:5482–5493.
-
- Stern C, Sommer S, Germann C, et al. Pelvic bone CT: can tin-filtered ultra-low-dose CT and virtual radiographs be used as alternative for standard CT and digital radiographs? Eur Radiol . 2021;31:6793–6801.
-
- Grunz JP, Heidenreich JF, Lennartz S, et al. Spectral shaping via tin prefiltration in ultra-high-resolution photon-counting and energy-integrating detector CT of the temporal bone. Invest Radiol . 2022;57:819–825.
-
- Rajendran K, Voss BA, Zhou W, et al. Dose reduction for sinus and temporal bone imaging using photon-counting detector CT with an additional tin filter. Invest Radiol . 2020;55:91–100.
-
- Hackenbroch C, Schüle S, Halt D, et al. Metal artifact reduction with tin prefiltration in computed tomography: a cadaver study for comparison with other novel techniques. Invest Radiol . 2022;57:194–203.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical
Research Materials