Radiation dose to multidisciplinary staff members during complex interventional procedures
- PMID: 38241981
- DOI: 10.1016/j.radi.2024.01.010
Radiation dose to multidisciplinary staff members during complex interventional procedures
Abstract
Introduction: Complex interventional radiology procedures involve extensive fluoroscopy and image acquisition while staff are in-room. Monitoring occupational radiation dose is crucial in optimization. The purpose was to determine radiation doses received by staff involved in complex interventional procedures performed in a dedicated vascular or neuro intervention room.
Methods: Individual real-time radiation dose for all staff involved in vascular and neuro-interventional procedures in adult patients was recorded over a one-year period using wireless electronic dosimeters attached to the apron thyroid shield. A reference dosimeter was attached to the C-arm near the tube housing to measure scattered, unshielded radiation. Radiology staff carried shoulder thermo-luminescent dosimeters with monthly read-out to monitor dose over time.
Results: Occupational radiation dose was measured in 99 interventional procedures. In many cases prostate artery embolization procedures exposed radiologists to high radiation doses with a median of 15.0 μSv and a very large spread, i.e. 0.2-152.5 μSv. In all procedures except uterine fibroid embolization radiographers were exposed to lower doses than those of radiologists, with endovascular aortic repair being the procedure with highest median exposure to assisting radiographers, i.e. 2.2 μSv ranging from 0.1 to 36.1 μSv. Median radiation dose for the reference dosimeter was 670 μGy while median staff dose for all procedures combined was 3.2 μGy.
Conclusion: Radiation doses for multiple staff were determined and the ratio between staff dose and reference dosimeter indicated proper use of shielding in general. Some high-dose procedures may need further optimization for certain staff members, especially those not primarily employed in radiology.
Implications for practice: The study provides benchmark doses that may be used widely in audits and in the ongoing effort to optimize radiation protection for staff in interventional radiology.
Keywords: Endovascular aortic repair; Interventional radiology; Prostate artery embolization; Radiation protection; Transjugular intrahepatic portosystemic shunt; Uterine fibroid embolization.
Copyright © 2024 The Author(s). Published by Elsevier Ltd.. All rights reserved.
Similar articles
-
Radiation exposure to operating room personnel and patients during endovascular procedures.J Vasc Surg. 2013 Sep;58(3):702-9. doi: 10.1016/j.jvs.2013.02.032. Epub 2013 Jun 28. J Vasc Surg. 2013. PMID: 23810300 Clinical Trial.
-
Evaluation of scattered radiation dose received by medical staff during uterine artery embolization in the operating room.Technol Health Care. 2020;28(S1):3-11. doi: 10.3233/THC-209002. Technol Health Care. 2020. PMID: 32364139 Free PMC article.
-
Dual fluoroscopy with live-image digital zooming significantly reduces patient and operating staff radiation during fenestrated-branched endovascular aortic aneurysm repair.J Vasc Surg. 2021 Feb;73(2):601-607. doi: 10.1016/j.jvs.2020.05.031. Epub 2020 May 27. J Vasc Surg. 2021. PMID: 32473339
-
Occupational radiation doses to operators performing fluoroscopically-guided procedures.Health Phys. 2012 Jul;103(1):80-99. doi: 10.1097/HP.0b013e31824dae76. Health Phys. 2012. PMID: 22647920 Free PMC article. Review.
-
Avoidance of radiation injuries from medical interventional procedures.Ann ICRP. 2000;30(2):7-67. doi: 10.1016/S0146-6453(01)00004-5. Ann ICRP. 2000. PMID: 11459599 Review.
Cited by
-
Exposure to ionizing radiation of medical staff performing vascular and interventional radiology procedures.Int J Occup Med Environ Health. 2024 Nov 6;37(4):403-410. doi: 10.13075/ijomeh.1896.02462. Epub 2024 Sep 11. Int J Occup Med Environ Health. 2024. PMID: 39291364 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources