Treatment of atrial fibrillation with second-generation cryoballoon followed by contact-sensing radiofrequency catheter ablation for arrhythmia recurrences-results of a 5-year follow-up
- PMID: 38261100
- DOI: 10.1007/s10840-024-01752-8
Treatment of atrial fibrillation with second-generation cryoballoon followed by contact-sensing radiofrequency catheter ablation for arrhythmia recurrences-results of a 5-year follow-up
Abstract
Introduction: The aim of this study was to report the long-term follow-up results of cryoballoon (CB) ablation in patients with atrial fibrillation.
Methods: All consecutive patients who underwent second-generation CB ablation from February 2015 to December 2017 were included in our study. In all procedures, we used a 28-mm CB placed via a single transseptal puncture guided by intracardiac ultrasound. A 20-mm octapolar intraluminal circular catheter was used for intracardiac recordings. A single 180-s freeze strategy was employed. Repeated procedures were performed with a 3D mapping system and radiofrequency catheters.
Results: A total of 126 patients (69.8% male, mean age 57 ± 11 years), of which 77.0% had paroxysmal atrial fibrillation (PAF), were included in the study. After a 5-year period, 52.4% of patients were in sinus rhythm without AF recurrence, off antiarrhythmic drugs. A total of 61.9% of patients were free of AF recurrence when redo PVI procedures were performed. When accounting for redo pulmonary vein isolation and antiarrhythmic drugs, a total of 73.8% of the patients were without AF recurrence in long-term follow-up. The patients who underwent redo pulmonary vein isolation procedures had statistically significant lower rates of AF recurrence (p = 0.006). In patients with PAF, long-term success rates improved from 62.9 to 79.4% for patients who underwent the redo procedure (p = 0.020). In patients with persistent atrial fibrillation (PersAF), success rates went up from 41.4 to 55.1% for patients with single or repeated PVI procedure (p = 0.071). In the whole cohort, a total of 3 (2.4%) procedure-related major complications occurred which included persistent PNP, arterial pseudoaneurysm, and arteriovenous fistula.
Conclusion: Our data suggest a favorable long-term safety and efficacy profile of second-generation CB ablation. In the mixed paroxysmal and persistent population, up to 73.8% of patients remained free of AF recurrence in the 5-year follow-up, when accounting for redo procedures and AADs. Only 2.4% of patients experienced major complications of the ablation procedure, none with permanent sequelae.
Keywords: Atrial fibrillation; Catheter ablation; Cryoballoon; Long-term follow-up.
© 2024. The Author(s), under exclusive licence to Springer Science+Business Media, LLC, part of Springer Nature.
References
-
- Metzner A, Reissmann B, Rausch P, Mathew S, Wohlmuth P, Tilz R, et al. One-year clinical outcome after pulmonary vein isolation using the second-generation 28-mm cryoballoon. Circ Arrhythm Electrophysiol. 2014;7(2):288–92. https://doi.org/10.1161/CIRCEP.114.001473 . - DOI - PubMed
-
- Wissner E, Heeger CH, Grahn H, Reissmann B, Wohlmuth P, Lemes C, et al. One-year clinical success of a ‘no-bonus’ freeze protocol using the second-generation 28 mm cryoballoon for pulmonary vein isolation. Europace. 2015;17(8):1236–40. https://doi.org/10.1093/europace/euv024 . - DOI - PubMed
-
- Heeger CH, Wissner E, Knöll M, Knoop B, Reissmann B, Mathew S, et al. Three-year clinical outcome after 2nd-generation cryoballoon-based pulmonary vein isolation for the treatment of paroxysmal and persistent atrial fibrillation - a 2-center experience. Circ J. 2017;81(7):974–80. https://doi.org/10.1253/circj.CJ-16-1334 . - DOI - PubMed
-
- Gökoğlan Y, Mohanty S, Güneş MF, Trivedi C, Santangeli P, Gianni C, et al. Pulmonary vein antrum isolation in patients with paroxysmal atrial fibrillation: more than a decade of follow-up. Circ Arrhythm Electrophysiol. 2016;9(5):e003660. https://doi.org/10.1161/CIRCEP.115.003660 . - DOI - PubMed
-
- Ouyang F, Tilz R, Chun J, Schmidt B, Wissner E, Zerm T, et al. Long-term results of catheter ablation in paroxysmal atrial fibrillation: lessons from a 5-year follow-up. Circulation. 2010;122(23):2368–77. https://doi.org/10.1161/CIRCULATIONAHA.110.946806 . - DOI - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Miscellaneous