. 2024 Feb 28;69(3):349-353.
doi: 10.4187/respcare.11280.

Simplified Mechanical Power Calculation in Patients Receiving Pressure-Regulated, Volume-Targeted Hybrid Modes of Ventilation

Affiliations

Simplified Mechanical Power Calculation in Patients Receiving Pressure-Regulated, Volume-Targeted Hybrid Modes of Ventilation

Abeer Santarisi et al. Respir Care. .
No abstract available

Keywords: adaptive pressure ventilation; adaptive support ventilation; mechanical power; pressure control-mechanical power; volume control-mechanical power.

PubMed Disclaimer

Conflict of interest statement

Drs Baedorf Kassis and Talmor disclose relationships with Hamilton Medical. The remaining authors have disclosed no conflicts of interest.

Figures

Fig. 1.
Fig. 1.
Univariate linear regression showing significant correlation between pressure controlled mechanical power (MP) and volume controlled MP (VC-MP) in adaptive pressure ventilation (panels A and B) and adaptive support ventilation (panels C and D) and waveform-MP with R2 of 0.93, 0.95, 0.81, and 0.94, respectively (P values were all < .001). The line of identity demonstrated consistently lower VC-MP measure than waveform-MP. MP = mechanical power; PC-MP = pressure controlled mechanical power; VC-MP = volume controlled mechanical power.
Fig. 2.
Fig. 2.
In Wilcoxon signed-rank test, there was no difference between waveform-mechanical power (MP) and pressure controlled MP values (adaptive pressure ventilation [APV] panels A and B: 28.2 [24.7–33.4] J/min, P = .94; and adaptive support ventilation [ASV] panels C and D: 28.5 [21.5–37.7] J/min, P = .10). A significant difference was detected between waveform-MP and volume controlled MP (APV: 22.3 [19.2–29.8] J/min, P < .001; and ASV: 21.8 [18.9–31.8] J/min, P < .001]. MP = mechanical power; PC-MP = pressure controlled mechanical power; VC-MP = volume controlled mechanical power.

References

    1. Gattinoni L, Tonetti T, Cressoni M, Cadringher P, Herrmann P, Moerer O, et al. . Ventilator-related causes of lung injury: the mechanical power. Intensive Care Med 2016;42(10):1567–1575. - PubMed
    1. Paudel R, Trinkle CA, Waters CM, Robinson LE, Cassity E, Sturgill JL, et al. . Mechanical power: a new concept in mechanical ventilation. Am J Med Sci 2021;362(6):537–545. - PMC - PubMed