. 2024 Jun;33(3):176-183.
doi: 10.1080/13645706.2024.2313032. Epub 2024 Feb 9.

Hybrid optical-vision tracking in laparoscopy: accuracy of navigation and ultrasound reconstruction

Affiliations
Free article

Hybrid optical-vision tracking in laparoscopy: accuracy of navigation and ultrasound reconstruction

Luca Boretto et al. Minim Invasive Ther Allied Technol. 2024 Jun.
Free article

Abstract

Introduction: The use of laparoscopic and robotic liver surgery is increasing. However, it presents challenges such as limited field of view and organ deformations. Surgeons rely on laparoscopic ultrasound (LUS) for guidance, but mentally correlating ultrasound images with pre-operative volumes can be difficult. In this direction, surgical navigation systems are being developed to assist with intra-operative understanding. One approach is performing intra-operative ultrasound 3D reconstructions. The accuracy of these reconstructions depends on tracking the LUS probe.

Material and methods: This study evaluates the accuracy of LUS probe tracking and ultrasound 3D reconstruction using a hybrid tracking approach. The LUS probe is tracked from laparoscope images, while an optical tracker tracks the laparoscope. The accuracy of hybrid tracking is compared to full optical tracking using a dual-modality tool. Ultrasound 3D reconstruction accuracy is assessed on an abdominal phantom with CT transformed into the optical tracker's coordinate system.

Results: Hybrid tracking achieves a tracking error < 2 mm within 10 cm between the laparoscope and the LUS probe. The ultrasound reconstruction accuracy is approximately 2 mm.

Conclusion: Hybrid tracking shows promising results that can meet the required navigation accuracy for laparoscopic liver surgery.

Keywords: Laparoscopic ultrasound tracking; laparoscopic liver surgery; surgical navigation; ultrasound reconstruction.

PubMed Disclaimer

MeSH terms

LinkOut - more resources