Thin slice photon-counting CT coronary angiography compared to conventional CT: Objective image quality and clinical radiation dose assessment
- PMID: 38358113
- DOI: 10.1002/mp.16992
Thin slice photon-counting CT coronary angiography compared to conventional CT: Objective image quality and clinical radiation dose assessment
Abstract
Background: Photon-counting CT (PCCT) is the next-generation CT scanner that enables improved spatial resolution and spectral imaging. For full spectral processing, higher tube voltages compared to conventional CT are necessary to achieve the required spectral separation. This generated interest in the potential influence of thin slice high tube voltage PCCT on overall image quality and consequently on radiation dose.
Purpose: This study first evaluated tube voltages and radiation doses applied in patients who underwent coronary CT angiography with PCCT and energy-integrating detector CT (EID-CT). Next, image quality of PCCT and EID-CT was objectively evaluated in a phantom study simulating different patient sizes at these tube voltages and radiation doses.
Methods: We conducted a retrospective analysis of clinical doses of patients scanned on a conventional and PCCT system. Average patient water equivalent diameters for different tube voltages were extracted from the dose reports for both EID-CT and PCCT. A conical phantom made of polyethylene with multiple diameters (26/31/36 cm) representing different patient sizes and containing an iodine insert was scanned with a EID-CT scanner using tube voltages and phantom diameters that match the patient scans and characteristics. Next, phantom scans were made with PCCT at a fixed tube voltage of 120 kV and with CTDIVOL values and phantom diameters identical to the EID-CT scans. Clinical image reconstructions at 0.6 mm slice thickness for conventional CT were compared to PCCT images with 0.4 mm slice thickness. Image quality was quantified using the detectability index (d'), which estimated the visibility of a 3 mm diameter contrast-enhanced coronary artery by considering noise, contrast, resolution, and human visual perception. Alongside d', noise, contrast and resolution were also individually assessed. In addition, the influence of various kernels (Bv40/Bv44/Bv48/Bv56), quantum iterative reconstruction strengths (QIR, 3/4) and monoenergetic levels (40/45/50/55 keV) for PCCT on d' was investigated.
Results: In this study, 143 patients were included: 47 were scanned on PCCT (120 kV) and the remaining on EID-CT (74 small-sized at 70 kV, 18 medium-sized at 80 kV and four large-sized at 90 kV). EID-CT showed 7%-17% higher d' than PCCT with Bv40 kernel and strength four for small/medium patients. Lower monoenergetic images (40 keV) helped mitigate the difference to 1%-6%. For large patients, PCCT's detectability was up to 31% higher than EID-CT. PCCT has thinner slices but similar noise levels for similar reconstruction parameters. The noise increased with lower keV levels in PCCT (≈30% increase), but higher QIR strengths reduced noise. PCCT's iodine contrast was stable across patient sizes, while EID-CT had 33% less contrast in large patients than in small-sized patients.
Conclusion: At 120 kV, thin slice PCCT enables CCTA in phantom scans representing large patients without raising radiation dose or affecting vessel detectability. However, higher doses are needed for small and medium-sized patients to obtain a similar image quality as in EID-CT. The alternative of using lower mono-energetic levels requires further evaluation in clinical practice.
Keywords: coronary computed tomography angiography; energy‐integrating computed tomography; image quality; photon‐counting computed tomography.
© 2024 The Authors. Medical Physics published by Wiley Periodicals LLC on behalf of American Association of Physicists in Medicine.
Similar articles
-
Photon-counting CT versus energy-integrating detectors for cardiac imaging: a systematic review of evidence from in vivo human studies on image quality and radiation dose.BMC Med Imaging. 2025 Jul 25;25(1):295. doi: 10.1186/s12880-025-01825-8. BMC Med Imaging. 2025. PMID: 40713483 Free PMC article.
-
A systematic task-based image quality assessment of photon-counting and energy integrating CT as a function of reconstruction kernel and phantom size.Med Phys. 2024 Feb;51(2):1047-1060. doi: 10.1002/mp.16619. Epub 2023 Jul 19. Med Phys. 2024. PMID: 37469179 Free PMC article.
-
Ultra-High-Resolution Coronary CT Angiography With Photon-Counting Detector CT: Feasibility and Image Characterization.Invest Radiol. 2022 Dec 1;57(12):780-788. doi: 10.1097/RLI.0000000000000897. Epub 2022 May 31. Invest Radiol. 2022. PMID: 35640019 Free PMC article.
-
Universal 120-kV Dual-Source Ultra-High Pitch Protocol on the Photon-Counting CT System for Pediatric Abdomen of All Sizes: A Phantom Investigation Comparing With Energy-Integrating CT.Invest Radiol. 2024 Oct 1;59(10):719-726. doi: 10.1097/RLI.0000000000001080. Epub 2024 Apr 10. Invest Radiol. 2024. PMID: 38595181
-
Photon-Counting Computed Tomography for Microstructural Imaging of Bone and Joints.Curr Osteoporos Rep. 2024 Aug;22(4):387-395. doi: 10.1007/s11914-024-00876-0. Epub 2024 Jun 4. Curr Osteoporos Rep. 2024. PMID: 38833188 Review.
Cited by
-
Impact of Vessel Size, Dose Levels, and Body Habitus on Iodine Quantification in Cardiovascular Photon-Counting Computed Tomography.Br J Radiol. 2025 Jul 30:tqaf181. doi: 10.1093/bjr/tqaf181. Online ahead of print. Br J Radiol. 2025. PMID: 40737438 Free PMC article.
-
Gaussian Function Model for Task-Specific Evaluation in Medical Imaging: A Theoretical Investigation.J Imaging Inform Med. 2025 Apr 24. doi: 10.1007/s10278-025-01511-9. Online ahead of print. J Imaging Inform Med. 2025. PMID: 40274693
-
Approaches, advantages, and challenges to photon counting detector and multi-energy CT.Abdom Radiol (NY). 2024 Sep;49(9):3251-3260. doi: 10.1007/s00261-024-04357-x. Epub 2024 May 15. Abdom Radiol (NY). 2024. PMID: 38744702 Review.
-
Photon-counting CT versus energy-integrating detectors for cardiac imaging: a systematic review of evidence from in vivo human studies on image quality and radiation dose.BMC Med Imaging. 2025 Jul 25;25(1):295. doi: 10.1186/s12880-025-01825-8. BMC Med Imaging. 2025. PMID: 40713483 Free PMC article.
References
REFERENCES
-
- Boccalini S, Si‐Mohamed SA, Lacombe H, et al. First in‐human results of computed tomography angiography for coronary stent assessment with a spectral photon counting computed tomography. Invest Radiol. 2022;57(4):212‐221. doi:10.1097/RLI.0000000000000835
-
- Si‐Mohamed SA, Boccalini S, Lacombe H, et al. Coronary CT angiography with photon‐counting CT: first‐in‐human results. Radiology. 2022;303(2):303‐313. doi:10.1148/radiol.211780
-
- Mergen V, Sartoretti T, Baer‐Beck M, et al. Ultra‐high‐resolution coronary CT angiography with photon‐counting detector CT: feasibility and image characterization. Invest Radiol. 2022;57(12):780‐788. doi:10.1097/rli.0000000000000897
-
- Symons R, Sandfort V, Mallek M, Ulzheimer S, Pourmorteza A. Coronary artery calcium scoring with photon‐counting CT: first in vivo human experience. Int J Cardiovasc Imaging. 2019;35(4):733‐739. doi:10.1007/s10554‐018‐1499‐6
-
- Dobrolinska MM, van der Werf NR, van der Bie J, et al. Radiation dose optimization for photon‐counting CT coronary artery calcium scoring for different patient sizes: a dynamic phantom study. Eur Radiol. 2023;33:4668‐4675. doi:10.1007/s00330‐023‐09434‐1
MeSH terms