Establishment of Diagnostic Reference Levels in Portuguese Interventional Radiology departments
- PMID: 38382425
- DOI: 10.1016/j.ejrad.2024.111377
Establishment of Diagnostic Reference Levels in Portuguese Interventional Radiology departments
Abstract
Purpose: To establish Portuguese Diagnostic Reference Levels (DRLs), for six body fluoroscopy guided interventional procedures (FGIP).
Method: A retrospective study was conducted in five interventional departments most representative of Interventional Radiology (IR) practice. Dose values, in terms of air kerma area product (PKA in Gy.cm2), air kerma at the patient entrance reference point (Ka,r in mGy), and exposure parameters (fluoroscopy time (FT) and number of cine runs) were collected. Examinations were selected per procedure (at least 20), according to the antero-posterior and lateral diameter mean value (±5 cm), measured on previous Computed Tomography (CT) examinations.
Results: Data of 489 body FGIP show a large variation on dose values per procedure and per department. National DRLs in terms of PKA were 20.2 Gy.cm2 for Percutaneous transhepatic biliary drainage (PTBD), 98.2 Gy.cm2 for Bronchial artery embolisation (BAE), 247.7 Gy.cm2 for Transarterial chemoembolisation (TACE), 331.6 Gy.cm2 for Inferior epigastric arteries embolisation (IEAE), 312.0 Gy.cm2 for Transjugular intrahepatic portosystemic shunt (TIPS) and 19.3 Gy.cm2 for Endovascular treatment of femoral popliteal arteries (ETFPA).
Conclusions: This is the first study reporting Interventional Radiology DRLs in Portugal and we propose preliminary national estimates for the six more common body FGIP. The results of this study will be presented and discussed with all Portuguese IR departments, to promote procedures optimisation.
Keywords: Diagnostic Reference Levels; Dose values; Interventional Radiology; Radiation Protection.
Copyright © 2024 Elsevier B.V. All rights reserved.
Conflict of interest statement
Declaration of competing interest The authors declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.
Similar articles
-
Institutional Diagnostic Reference Levels and Peak Skin Doses in selected diagnostic and therapeutic interventional radiology procedures.Phys Med. 2021 Sep;89:63-71. doi: 10.1016/j.ejmp.2021.07.029. Epub 2021 Aug 2. Phys Med. 2021. PMID: 34352677
-
DIAGNOSTIC REFERENCE LEVELS AND COMPLEXITY INDICES IN INTERVENTIONAL RADIOLOGY.Radiat Prot Dosimetry. 2023 Mar 2;199(3):ncac267 254 261-261. doi: 10.1093/rpd/ncac267. Radiat Prot Dosimetry. 2023. PMID: 36566495
-
Patient Radiation Doses in IR Procedures: The American College of Radiology Dose Index Registry-Fluoroscopy Pilot.J Vasc Interv Radiol. 2023 Apr;34(4):544-555.e11. doi: 10.1016/j.jvir.2022.11.003. Epub 2022 Nov 13. J Vasc Interv Radiol. 2023. PMID: 36379286
-
Notifications and alerts in patient dose values for computed tomography and fluoroscopy-guided interventional procedures.Eur Radiol. 2022 Aug;32(8):5525-5531. doi: 10.1007/s00330-022-08675-w. Epub 2022 Mar 16. Eur Radiol. 2022. PMID: 35294584 Free PMC article. Review.
-
Review of air kerma-area product, effective dose and dose conversion coefficients for non-cardiac interventional fluoroscopy procedures.Med Phys. 2020 Mar;47(3):975-982. doi: 10.1002/mp.13990. Epub 2020 Jan 20. Med Phys. 2020. PMID: 31880815 Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous