Intraocular lens damage associated with posterior capsulotomy: a comparison of intraocular lens designs and four different Nd:YAG laser instruments
- PMID: 3841118
- DOI: 10.1016/s0146-2776(85)80133-6
Intraocular lens damage associated with posterior capsulotomy: a comparison of intraocular lens designs and four different Nd:YAG laser instruments
Abstract
A comparison of intraocular lens damage associated with four different Nd:YAG laser instruments was performed using polarized light, scanning electron microscopy, and a qualitative measurement of optical resolution. Five intraocular lens designs, each of which provided a different separation between the intraocular lens and a model of the posterior capsule, were tested with each laser instrument. There was no statistically significant difference in the frequency of lens pitting caused by each laser. Intraocular lenses designed with a theoretical separation of 0.25 mm or more between the lens and the posterior capsule sustained no damage with any of the laser instruments used in this study.
Similar articles
-
Neodymium: YAG laser capsulotomy with a biconvex intraocular lens.J Cataract Refract Surg. 1989 Jul;15(4):435-6. doi: 10.1016/s0886-3350(89)80065-3. J Cataract Refract Surg. 1989. PMID: 2778690
-
Persistent elevation in intraocular pressure after Nd:YAG laser treatment.Ophthalmic Surg. 1986 Aug;17(8):465-6. Ophthalmic Surg. 1986. PMID: 3748536
-
Spectrum of Nd:YAG laser-induced intraocular lens damage in explanted lenses.J Cataract Refract Surg. 1990 Jul;16(4):495-500. doi: 10.1016/s0886-3350(13)80806-1. J Cataract Refract Surg. 1990. PMID: 2380933
-
Review of the clinical literature on the use of the Nd:YAG laser for posterior capsulotomy.J Cataract Refract Surg. 1986 Mar;12(2):162-70. doi: 10.1016/s0886-3350(86)80034-7. J Cataract Refract Surg. 1986. PMID: 3517297 Review.
-
Efficacy of different intraocular lens materials and optic edge designs in preventing posterior capsular opacification: a meta-analysis.Am J Ophthalmol. 2007 Mar;143(3):428-36. doi: 10.1016/j.ajo.2006.11.045. Epub 2006 Dec 28. Am J Ophthalmol. 2007. PMID: 17224119
Cited by
-
Optical scattering measurements of laser induced damage in the intraocular lens.PLoS One. 2012;7(2):e31764. doi: 10.1371/journal.pone.0031764. Epub 2012 Feb 10. PLoS One. 2012. PMID: 22348128 Free PMC article.