An Adaptive Harmonic Separation Technique for Ultrasound Harmonic Imaging
- PMID: 38413294
- DOI: 10.1016/j.ultrasmedbio.2024.02.001
An Adaptive Harmonic Separation Technique for Ultrasound Harmonic Imaging
Abstract
Objective: An adaptive harmonic separation (HS) technique is proposed to overcome the limitations in conventional filtering techniques for ultrasound (US) tissue harmonic imaging (THI).
Methods: Based on expectation-maximization source separation, the proposed HS technique adaptively models the depth-varying fundamental and harmonic components in the frequency domain and separates the two by applying their calculated posterior probabilities. Phantom experiments with a Tx center frequency of 2 MHz are conducted to evaluate the proposed HS-based US THI schemes.
Results: The phantom images show that the proposed single-pulse THI scheme utilizing the HS technique provides not only an average improvement of 19.2% in axial resolution compared to the conventional bandpass filtering scheme but also similar image quality to that of the conventional pulse-inversion (PI) scheme which requires two Tx/Rx sequences for each scan line. Furthermore, when combined with the PI technique, the HS technique provides a uniform axial resolution over the entire 170 mm imaging depth with an average improvement of 17.1% compared to the conventional PI scheme.
Conclusion: These results show that the proposed adaptive HS technique is capable of improving both the frame rate and the image quality of US THI.
Keywords: Expectation-maximization algorithm; Pulse inversion; Source separation; Ultrasound harmonic imaging.
Copyright © 2024. Published by Elsevier Inc.
Conflict of interest statement
Conflict of interest disclosure Hansol Yoon and Tai-Kyong Song declare that they have no conflict of interest.
Similar articles
-
Adaptive Ultrasound Tissue Harmonic Imaging Based on an Improved Ensemble Empirical Mode Decomposition Algorithm.Ultrason Imaging. 2020 Mar;42(2):57-73. doi: 10.1177/0161734619900147. Epub 2020 Jan 29. Ultrason Imaging. 2020. PMID: 31994455
-
An Adaptive Synthetic Aperture Method Applied to Ultrasound Tissue Harmonic Imaging.IEEE Trans Ultrason Ferroelectr Freq Control. 2018 Apr;65(4):557-569. doi: 10.1109/TUFFC.2018.2799870. IEEE Trans Ultrason Ferroelectr Freq Control. 2018. PMID: 29610086
-
Intravascular Ultrasound Transducer by Using Polarization Inversion Technique for Tissue Harmonic Imaging: Modeling and Experiments.IEEE Trans Biomed Eng. 2020 Dec;67(12):3380-3391. doi: 10.1109/TBME.2020.2986284. Epub 2020 Nov 19. IEEE Trans Biomed Eng. 2020. PMID: 32286955
-
A Primer on the Physical Principles of Tissue Harmonic Imaging.Radiographics. 2015 Nov-Dec;35(7):1955-64. doi: 10.1148/rg.2015140338. Radiographics. 2015. PMID: 26562232 Review.
-
Broadband reduction of the second harmonic distortion during nonlinear ultrasound wave propagation.Ultrasound Med Biol. 2010 Oct;36(10):1568-80. doi: 10.1016/j.ultrasmedbio.2010.06.006. Ultrasound Med Biol. 2010. PMID: 20800962 Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Research Materials
Miscellaneous