Effect of NaOCl on cyclic fatigue resistance
- PMID: 38415037
- PMCID: PMC10895534
- DOI: 10.6026/973206300191348
Effect of NaOCl on cyclic fatigue resistance
Abstract
The impact of 5.25 percent NaOCL on the resistance offered by TruNatomy, Hero Gold and ProtaperGold against cyclic fracture as against the treatment of NiTi rotary files with distilled water is of interest to dentists. Inside the stainless steel blocks, man-made canals were created for the purpose of testing the cyclic fatigue. 60 endodontic files were taken as study specimens. It was observed that values of number of cycles to fracture in NaOCl among three file systems were in the order of TruNatomy (1053.50 ± 134.81)>Hero Gold (652.66 ± 58.66) > ProtaperGold (494.50 ± 47.69). The TruNatomy file system reflected greatest cyclic fatigue resistance. It was also found that cyclic fatigue resistance in NiTi rotary files studied here is not hampered by 5.25% NaOCl.
Keywords: Cyclic fatigue; HeroGold; Protaper Gold; TruNatomy; resistance to cyclic fracture.
© 2023 Biomedical Informatics.
Similar articles
-
Comparing the EdgeFile X3, GenEndo, HeroGold, and ProTaper Gold Rotary Instruments Regarding the Effect of Different Concentrations and Temperature of NaOCl on Cyclic Fatigue Resistance.J Contemp Dent Pract. 2023 Sep 1;24(9):715-721. doi: 10.5005/jp-journals-10024-3515. J Contemp Dent Pract. 2023. PMID: 38152947
-
Comparison of Cyclic Fatigue Resistance of Novel TruNatomy Files with Conventional Endodontic Files: An In Vitro SEM Study.J Contemp Dent Pract. 2021 Nov 1;22(11):1243-1249. J Contemp Dent Pract. 2021. PMID: 35343448
-
In Vitro Comparison of Cyclic Fatigue Resistance of ProTaper Next and TruNatomy Endodontic Rotary Files.J Pharm Bioallied Sci. 2024 Jul;16(Suppl 3):S2126-S2128. doi: 10.4103/jpbs.jpbs_107_24. Epub 2024 Apr 3. J Pharm Bioallied Sci. 2024. PMID: 39346307 Free PMC article.
-
Cyclic Fatigue Resistance of Rotary versus Reciprocating Endodontic Files: A Systematic Review and Meta-Analysis.J Clin Med. 2024 Feb 2;13(3):882. doi: 10.3390/jcm13030882. J Clin Med. 2024. PMID: 38337577 Free PMC article. Review.
-
Cyclic Fatigue Resistance of Glide Path Rotary Files: A Systematic Review of in Vitro Studies.Materials (Basel). 2022 Sep 26;15(19):6662. doi: 10.3390/ma15196662. Materials (Basel). 2022. PMID: 36234003 Free PMC article. Review.
Cited by
-
In Vitro Evaluation of Microbial Leakage in Different Root Canal Obturation Techniques.J Pharm Bioallied Sci. 2025 Jun;17(Suppl 2):S1220-S1222. doi: 10.4103/jpbs.jpbs_1649_24. Epub 2025 Jun 18. J Pharm Bioallied Sci. 2025. PMID: 40655617 Free PMC article.
References
LinkOut - more resources