Comparative evaluation of Vitek®2 and broth microdilution method for colistin susceptibility testing of Gram-negative isolates from intensive care unit in a tertiary care hospital
- PMID: 38447856
- DOI: 10.1016/j.ijmmb.2024.100559
Comparative evaluation of Vitek®2 and broth microdilution method for colistin susceptibility testing of Gram-negative isolates from intensive care unit in a tertiary care hospital
Abstract
Introduction: Colistin is the last resort treatment against resistant Gram-negative bacteria, necessitating reliable and rapid means for sensitivity testing of colistin. Automated systems like VITEK®2 are adopted to determine the minimum inhibitory concentration (MIC) due to easy usage. Broth microdilution (BMD) for colistin MIC was suggested by EUCAST and CLSI.
Objective: To compare and evaluate colistin MIC by BMD and VITEK®2 against Gram-negative organisms from the ICU in a tertiary care hospital.
Method: Clinically significant organisms isolated from ICU patients were included. MIC was determined using BMD and VITEK®2. Very major error (VME), major error (ME), essential agreement (EA), categorical agreement (CA), positive predictive value (PPV), negative predictive value (NPV), sensitivity, and specificity were analysed.
Result: 533 isolates were obtained from blood (435,81.60%), respiratory samples (57,10.70%), pus and exudates (20,3.80%), urine (18,3.40%), and CSF (3,0.60%). The Enterobacterales were K. pneumoniae (185,34.70%) E. coli (73,13.70%) and E. cloacae (26,4.90%) while non-fermenters were A. baumannii (209,39.20%) and P. aeruginosa (40,7.50%). The VITEK®2 sensitivity was >99%; specificity ranged from 14.28 to 52.94%. PPV was 93.81% while NPV was 93.75%. VME ranged from 47 to 100% between isolates. ME was up to 20%. The highest VME was obtained in E. coli (100%). The total EA and CA observed were 68.5% and 99.79% respectively.
Conclusion: Automated system VITEK®2 failed to detect the resistance in 32 (60%) isolates. The obtained VME and ME values were >3%, which is unacceptable as per the standard guidelines. EA of ≥90% wasn't obtained. Sensitivity for VITEK®2 was >99%, but had low specificity (14.28%). Hence, VITEK®2 is not reliable for colistin susceptibility testing.
Keywords: Broth microdilution; Colistin; Gram-negative bacteria; Minimum inhibitory concentration; VITEK®2.
Copyright © 2024 The Authors. Published by Elsevier B.V. All rights reserved.
Conflict of interest statement
Declaration of competing interest The authors declare the following financial interests/personal relationships which may be considered as potential competing interests: Dr. Chiranjay Mukhopadhyay reports financial support was provided by Centers for Disease Control and Prevention. If there are other authors, they declare that they have no known competing financial interests or personal relationships that could have appeared to influence the work reported in this paper.