. 2024 Jun;69(6):1956-1962.
doi: 10.1007/s10620-024-08305-z. Epub 2024 Mar 11.

Assessment of the Learning Curve for a Single-Use Disposable Duodenoscope

Affiliations

Assessment of the Learning Curve for a Single-Use Disposable Duodenoscope

Nasir Saleem et al. Dig Dis Sci. 2024 Jun.

Abstract

Background and aims: In response to documented duodenoscope-related infectious outbreaks of multidrug-resistant organisms, the Food and Drug Administration has recommended a transition to duodenoscopes with innovative designs, including duodenoscopes with disposable components or fully disposable duodenoscopes. We aim to characterize the learning curve (LC) for a single-use disposable duodenoscope.

Methods: We performed a retrospective analysis of a prospectively collected database from 31 patients who underwent ERCP by a single, experienced operator using the EXALT Model D® (Boston Scientific, Marlborough) disposable duodenoscope at a single tertiary referral center. The LC for this device was described by the number of cases needed to achieve proficiency using cumulative sum (CUSUM) analysis. Number of attempts to cannulate and time to cannulate the desired duct were assessed as separate endpoints. The overall mean number of attempts and overall mean time to cannulation were used as the target values in the respective CUSUM analyses. Proficiency was defined as the number of procedures where an inflection point was reached in the CUSUM graph. This observation indicates improving operator performance as shown by a decrease in the number of attempts and shortening of cannulation time after the defined number of procedures.

Results: Overall, 31 patients underwent ERCP using the EXALT Model D disposable duodenoscope by a single experienced endoscopist. 6 (19%) patients had a native papilla and the majority of these procedures were classified as ASGE complexity level 2 or above. The procedure was completed using solely the disposable duodenoscope in 27 patients (87%), while a reusable duodenoscope was required for procedure completion in 4 patients (13%). The cross-overs were distributed evenly across the performance period. Procedure-related adverse events included: post-ERCP pancreatitis (3%), bleeding (3%) and no perforations. In the analyses of both endpoints, an inflection of the CUSUM curves is achieved at 10 cases, indicating sustained reduction of cannulation attempts and time to cannulation.

Conclusion: Among experienced pancreaticobiliary endoscopists, approximately 10 ERCPs is the threshold whereby procedure-related factors including cannulation success and procedural time improves. Procedure-related adverse events are consistent with those expected with reusable duodenoscopes. The need to cross-over from single-use duodenoscope to reusable duodenoscope did not appear to be related to the learning curve, as they were evenly distributed across the study period. These results can be used to guide adoption of single-use duodenoscopes into clinical practice.

Keywords: Disposable duodenoscopes; ERCP; Learning curve.

PubMed Disclaimer

References

    1. U.S. Food and Drug Administration. https://www.fda.gov/medical-devices/reprocessing-reusable-medical-device... . Accessed August 14, 2022.
    1. Peery AF, Crockett SD, Murphy CC, Lund JL, Dellon ES, Williams JL et al. Burden and cost of gastrointestinal, liver, and pancreatic diseases in the United States: update 2018. Gastroenterology. 2019;156:254–72.e11. https://doi.org/10.1053/j.gastro.2018.08.063 . - DOI - PubMed
    1. Kovaleva J, Peters FT, van der Mei HC, Degener JE. Transmission of infection by flexible gastrointestinal endoscopy and bronchoscopy. Clin Microbiol Rev. 2013;26:231–254. https://doi.org/10.1128/CMR.00085-12 . - DOI - PubMed - PMC
    1. Rubin ZA, Murthy RK. Outbreaks associated with duodenoscopes: new challenges and controversies. Curr Opin Infect Dis. 2016;29:407–414. https://doi.org/10.1097/QCO.0000000000000290 . - DOI - PubMed
    1. Epstein L, Hunter JC, Arwady MA, Tsai V, Stein L, Gribogiannis M et al. New Delhi metallo-beta-lactamase-producing carbapenem-resistant Escherichia coli associated with exposure to duodenoscopes. JAMA. 2014;312:1447–1455. https://doi.org/10.1001/jama.2014.12720 . - DOI - PubMed - PMC