A device-related fistula between coronary artery and left atrial appendage following left atrial appendage closure: Case presentation
- PMID: 38468182
- DOI: 10.1111/jce.16224
A device-related fistula between coronary artery and left atrial appendage following left atrial appendage closure: Case presentation
Abstract
Introduction: Left atrial appendage (LAA) closure (LAAC) is considered a viable alternative to anticoagulation therapy for stroke prevention in nonvalvular atrial fibrillation, we report a case with a less common shunt resulting from a device-related coronary artery-appendage fistula (CAAF) following LAAC.
Methods and results: A 67-year-old male with a history of LAAC was referred to our emergency room with recurrent chest pain and palpitations and was diagnosed with ischemic angina pectoris. Subsequent coronary angiography (CAG) revealed 70% in-stent restenosis and an abnormal shunt of contrast originating from the left circumflex artery (LCA) to the LAA tip which did not exist before. The restenosis was successfully dilated using a drug-coated balloon, the procedure was safely completed without pericardial effusion. The patient had been implanted with a LAmbre occluder (Lifetech Scientific Corp.) in the previous LAAC procedure. This occluder had a lobe-disk design, and the distal umbrella was not fully opened after release, particularly in the lower portion. This could make the hooks embedded on the umbrella contact the LAA wall more tightly, possibly resulting in microperforation and coincidental impingement of the LCA. The epicardial adipose and hyperplastic tissue then chronically wrapped the perforated site, prevented blood outflow into the epicardium, and ultimately formed a CAAF.
Conclusion: CAAF is a rare complication after LAAC but may be underestimated, especially for lobe-disk designed occluders. Therefore, CAG is perhaps necessary to detect this complication.
Keywords: complication; coronary artery; fistula; left atrial appendage; occluder; shunt.
© 2024 Wiley Periodicals LLC.
References
REFERENCES
-
- Reddy VY, Doshi SK, Kar S, et al. 5‐year outcomes after left atrial appendage closure. J Am Coll Cardiol. 2017;70(24):2964‐2975.
-
- Orban M, Brunner S, Massberg S, Bauer A. Uncommon cause of dyspnoea after percutaneous closure of the left atrial appendage. Eur Heart J. 2017;38(38):2912.
-
- Beigel R, Wunderlich NC, Ho SY, Arsanjani R, Siegel RJ. The left atrial appendage: anatomy, function, and noninvasive evaluation. JACC: Cardiovasc Imaging. 2014;7(12):1251‐1265.
-
- Yu Y, Chu H, Wang B, He B, Fu G. Pericardial effusion following percutaneous left atrial appendage closure using the LAmbre device. Front Cardiovasc Med. 2023;10:10.
-
- Mogalapalli A, Kumar S, Lobo T, et al. Delayed pericardial effusion following left atrial appendage closure: a 5‐year single‐center experience. J Invasive Cardiol. 2023;35(1):1.