Pre-excitation gradients for eddy current nulled convex optimized diffusion encoding (Pre-ENCODE)
- PMID: 38501914
- PMCID: PMC11142872
- DOI: 10.1002/mrm.30068
Pre-excitation gradients for eddy current nulled convex optimized diffusion encoding (Pre-ENCODE)
Abstract
Purpose: To evaluate the use of pre-excitation gradients for eddy current-nulled convex optimized diffusion encoding (Pre-ENCODE) to mitigate eddy current-induced image distortions in diffusion-weighted MRI (DWI).
Methods: DWI sequences using monopolar (MONO), ENCODE, and Pre-ENCODE were evaluated in terms of the minimum achievable echo time (TE ) and eddy current-induced image distortions using simulations, phantom experiments, and in vivo DWI in volunteers ( ).
Results: Pre-ENCODE provided a shorter TE than MONO (71.0 17.7ms vs. 77.6 22.9ms) and ENCODE (71.0 17.7ms vs. 86.2 14.2ms) in 100 of the simulated cases for a commercial 3T MRI system with b-values ranging from 500 to 3000 s/mm and in-plane spatial resolutions ranging from 1.0 to 3.0mm . Image distortion was estimated by intravoxel signal variance between diffusion encoding directions near the phantom edges and was significantly lower with Pre-ENCODE than with MONO (10.1 vs. 22.7 , ) and comparable to ENCODE (10.1 vs. 10.4 , ). In vivo measurements of apparent diffusion coefficients were similar in global brain pixels (0.37 [0.28,1.45] mm /s vs. 0.38 [0.28,1.45] mm /s, ) and increased in edge brain pixels (0.80 [0.17,1.49] mm /s vs. 0.70 [0.18,1.48] mm /s, ) for MONO compared to Pre-ENCODE.
Conclusion: Pre-ENCODE mitigated eddy current-induced image distortions for diffusion imaging with a shorter TE than MONO and ENCODE.
Keywords: diffusion; eddy currents; time‐optimal.
© 2024 International Society for Magnetic Resonance in Medicine.
Conflict of interest statement
Conflict of interest
The authors declare no potential conflict of interests.
Figures
References
-
- Fischer H, Ladebeck R. Echo-planar imaging image artifacts Echo-Planar Imaging ed F Schmitt, M Stehling and R Turner. 1998.
-
- Block W, Pauly J, Kerr AB, Nishimura D. Consistent fat suppression with compensated spectral-spatial pulses. Magn Reson Med. 1997;38(2):198–206. - PubMed
-
- Zur Y Design of improved spectral-spatial pulses for routine clinical use. Magn Reson Med. 2000;43(3):410–420. - PubMed
-
- Haselgrove JC, Moore JR. Correction for distortion of echo-planar images used to calculate the apparent diffusion coefficient. Magn Reson Med. 1996;36(6):960–964. - PubMed
-
- Lingamneni A, Hardy PA, Powell KA, Pelc NJ, White RD. Validation of cine phase-contrast MR imaging for motion analysis. J Magn Reson. 1995;5(3):331–338. - PubMed
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources