The reliability and validity of a novel wearable inertial sensor to measure the cervical proprioception

Affiliations

The reliability and validity of a novel wearable inertial sensor to measure the cervical proprioception

Jialin Wang et al. Med Eng Phys. 2024 Mar.

Abstract

Background: Proprioceptive function assessment is crucial in clinical practice for patients with chronic non-specific neck pain (CNNP) as it is a major issue affecting their condition.

Purpose: To verify the reliability and validity of baiobit sensor in measuring the neck proprioceptive function of CNNP patients.

Methods: Fifty-three CNNP patients were recruited (36 females, 17 males; age range 21-60 years) and were assessed for cervical joint position error by two blinded raters using the Baiobit sensor and laser pointer devices. The second measurement was conducted by the same rater 48 h later. Intra and inter-rater reliability of the Baiobit sensor was evaluated using the intra-class correlation coefficient (ICC), while the validity of the Baiobit sensor was established using the Spearman correlation coefficient.

Results: The Baiobit sensor demonstrated moderate to excellent intra-rater reliability in flexion, extension, left lateral flexion, right lateral flexion, and right rotation (ICCs=0.71∼0.85, 95 %CIs: 0.50∼0.91), left-rotation shows poor to good intra-rater reliability (ICC=0.56, 95 %CI: 0.25∼0.75). The Baiobit sensor also demonstrated moderate to excellent inter-rater reliability in flexion, extension, left lateral flexion, right lateral flexion, and right rotation (ICCs=0.80∼0.88, 95 %CIs: 0.65∼0.91), left-rotation shows poor to good intra-rater reliability (ICC=0.59, 95 %CI: 0.29∼0.76). Validity analysis showed that the Baiobit sensor had a range of low to high validity (r = 0.46∼0.88) for measuring cervical proprioception function, with lower validity observed in the left flexion direction. The Baiobit showed good absolute reliability with low SEM and MDC90 values (0.35°∼2.42°).

Conclusion: The new device could be used as an alternative tool to evaluate neck proprioception.

Keywords: Baiobit; Measurement; Neck pain; Proprioceptive; Sensor.