The reliability and validity of a novel wearable inertial sensor to measure the cervical proprioception
- PMID: 38508802
- DOI: 10.1016/j.medengphy.2024.104125
The reliability and validity of a novel wearable inertial sensor to measure the cervical proprioception
Abstract
Background: Proprioceptive function assessment is crucial in clinical practice for patients with chronic non-specific neck pain (CNNP) as it is a major issue affecting their condition.
Purpose: To verify the reliability and validity of baiobit sensor in measuring the neck proprioceptive function of CNNP patients.
Methods: Fifty-three CNNP patients were recruited (36 females, 17 males; age range 21-60 years) and were assessed for cervical joint position error by two blinded raters using the Baiobit sensor and laser pointer devices. The second measurement was conducted by the same rater 48 h later. Intra and inter-rater reliability of the Baiobit sensor was evaluated using the intra-class correlation coefficient (ICC), while the validity of the Baiobit sensor was established using the Spearman correlation coefficient.
Results: The Baiobit sensor demonstrated moderate to excellent intra-rater reliability in flexion, extension, left lateral flexion, right lateral flexion, and right rotation (ICCs=0.71∼0.85, 95 %CIs: 0.50∼0.91), left-rotation shows poor to good intra-rater reliability (ICC=0.56, 95 %CI: 0.25∼0.75). The Baiobit sensor also demonstrated moderate to excellent inter-rater reliability in flexion, extension, left lateral flexion, right lateral flexion, and right rotation (ICCs=0.80∼0.88, 95 %CIs: 0.65∼0.91), left-rotation shows poor to good intra-rater reliability (ICC=0.59, 95 %CI: 0.29∼0.76). Validity analysis showed that the Baiobit sensor had a range of low to high validity (r = 0.46∼0.88) for measuring cervical proprioception function, with lower validity observed in the left flexion direction. The Baiobit showed good absolute reliability with low SEM and MDC90 values (0.35°∼2.42°).
Conclusion: The new device could be used as an alternative tool to evaluate neck proprioception.
Keywords: Baiobit; Measurement; Neck pain; Proprioceptive; Sensor.
Published by Elsevier Ltd.
Conflict of interest statement
Declaration of competing interest The authors declare no conflict of interest.
Similar articles
-
Measurement properties of cervical joint position error in people with and without chronic neck pain.PLoS One. 2023 Oct 12;18(10):e0292798. doi: 10.1371/journal.pone.0292798. eCollection 2023. PLoS One. 2023. PMID: 37824590 Free PMC article.
-
Reliability and Validity of A Novel Device for Evaluating the Cervical Proprioception.Pain Ther. 2023 Jun;12(3):671-682. doi: 10.1007/s40122-023-00487-0. Epub 2023 Mar 3. Pain Ther. 2023. PMID: 36867364 Free PMC article.
-
Reliability of the Moover® 3D Inertial Motion Sensor in Greek Patients With Chronic Neck Pain in a Primary Care Urban Setting.Cureus. 2024 Aug 6;16(8):e66336. doi: 10.7759/cureus.66336. eCollection 2024 Aug. Cureus. 2024. PMID: 39246866 Free PMC article.
-
A new iPhone application for measuring active craniocervical range of motion in patients with non-specific neck pain: a reliability and validity study.Spine J. 2018 Mar;18(3):447-457. doi: 10.1016/j.spinee.2017.08.229. Epub 2017 Sep 7. Spine J. 2018. PMID: 28890223
-
Concurrent validity and reliability of measuring range of motion during the cervical flexion rotation test with a novel digital goniometer.BMC Musculoskelet Disord. 2020 Aug 11;21(1):535. doi: 10.1186/s12891-020-03525-6. BMC Musculoskelet Disord. 2020. PMID: 32781990 Free PMC article.
Cited by
-
Pelvic Control Characteristics During Static Balance in Patients with Multiple Sclerosis: A Novel Sensor-Based Study.J Clin Med. 2025 May 30;14(11):3854. doi: 10.3390/jcm14113854. J Clin Med. 2025. PMID: 40507619 Free PMC article.
-
Reliability and Validity of the KFORCE Sens® Inertial Sensor for Measuring Cervical Spine Proprioception in Patients with Non-Specific Chronic Neck Pain.Brain Sci. 2024 Nov 22;14(12):1165. doi: 10.3390/brainsci14121165. Brain Sci. 2024. PMID: 39766364 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical