Multicenter Study
. 2024 Oct;67(7):1539-1545.
doi: 10.1007/s10840-024-01791-1. Epub 2024 Mar 20.

Early real-world implant experience with a helix-fixation ventricular leadless pacemaker

Affiliations
Multicenter Study

Early real-world implant experience with a helix-fixation ventricular leadless pacemaker

Devi G Nair et al. J Interv Card Electrophysiol. 2024 Oct.

Abstract

Background: Roughly one in six patients receiving conventional transvenous pacemaker systems experience significant complications within 1 year of implant, mainly due to the transvenous lead and subcutaneous pocket. A new helix-fixation single-chamber ventricular leadless pacemaker (LP) system capable of pre-deployment exploratory electrical mapping is commercially available. Such an LP may mitigate complications while streamlining the implantation. In this study, the initial real-world implant experience of the helix-fixation LP was evaluated following its commercial release.

Methods: In patients indicated for single-chamber right ventricular pacing, helix-fixation Aveir VR LPs (Abbott, Abbott Park, IL) were implanted using the dedicated loading tool, introducer, and delivery catheter. Implant procedural characteristics, electrical parameters, and any 30-day procedure-related adverse events of consecutive implant attempts were retrospectively evaluated.

Results: A total of 167 patients with Class I indication for permanent pacing received implants in four North American centers (57% male, 70 years old). Pre-fixation electrical mapping of potential sites allowed repositioning to be avoided in 95.7% of patients. Median [interquartile range] LP procedure and fluoroscopy durations were 25.5 min [20.0, 35.0] and 5.7 min [4.0, 9.2], respectively. Pacing capture threshold, sensed R-wave amplitude, and impedance were 0.8 V [0.5, 1.3], 9.0 mV [6.0, 12.0], and 705 Ω [550, 910], respectively. Implantation was successful in 98.8% of patients, with 98.2% free from acute adverse events.

Conclusions: The initial, real-world experience of the helix-fixation ventricular leadless pacemaker demonstrated safe and efficient implantation with minimal repositioning, viable electrical metrics, and limited acute complications.

Keywords: Aveir; Helix fixation; Implant procedure; Leadless pacemaker; Pacemaker complications; Single-chamber pacemaker.

PubMed Disclaimer

References

    1. Cantillon DJ, et al. Complications and health care costs associated with transvenous cardiac pacemakers in a nationwide assessment. JACC: Clinical Electrophysiology. 2017;3(11):1296–305. - PubMed
    1. Kirkfeldt RE, Johansen JB, Nohr EA, Jorgensen OD, Nielsen JC. Complications after cardiac implantable electronic device implantations: an analysis of a complete, nationwide cohort in Denmark. Eur Heart J. 2014;35(18):1186–94. - DOI - PubMed
    1. Udo EO, et al. Incidence and predictors of short- and long-term complications in pacemaker therapy: the FOLLOWPACE study. Heart Rhythm. 2012;9(5):728–35. - DOI - PubMed
    1. Tobin K, Stewart J, Westveer D, Frumin H. Acute complications of permanent pacemaker implantation: their financial implication and relation to volume and operator experience. Am J Cardiol. 2000;85(6):774–6. - DOI - PubMed
    1. Sattar Y, et al. Complications of leadless vs conventional (lead) artificial pacemakers – a retrospective review. J Community Hosp Intern Med Perspect. 2020;10(4):328–33. - DOI - PubMed - PMC

Publication types

LinkOut - more resources