First in-human clinical performance of a new non-cavitating handheld lensectomy system in 665 consecutive cataract surgeries

Affiliations

First in-human clinical performance of a new non-cavitating handheld lensectomy system in 665 consecutive cataract surgeries

Tsontcho Ianchulev et al. J Cataract Refract Surg. .

Abstract

Purpose: To investigate the intraoperative performance and lens fragmentation efficacy of a non-cavitating handheld lensectomy system in mild, moderate, and severe cataract.

Setting: Ambulatory surgical centers.

Design: Retrospective consecutive case series.

Methods: 665 consecutive eyes underwent cataract surgery by 12 surgeons using a new handheld non-cavitating lensectomy system for nuclear fragmentations and extraction. Intraoperative measurements included surgical time, miLOOP pretreatment, and irrigation fluid use.

Results: Of the 665 eyes, 38 (6%), 468 (70%), 126 (19%), and 33 (5%) were of grade 1, 2, 3, and 4 nuclear densities, respectively, as graded by the surgeon intraoperatively. Successful nuclear fragmentation, lens extraction, and cortical removal were achieved in all eyes. Total nucleus fragmentation and extraction times were 70.1 seconds, 100.3 seconds, 132.6 seconds, and 287.9 seconds for grades 1, 2, 3, and 4, respectively ( P < .001). In addition, irrigation and aspiration cortical removal times were 64.1 seconds, 51.1 seconds, 48.5 seconds, and 59.0 seconds, respectively ( P = .14). There was a low rate of capsular tear (3 cases in 665 surgeries, 0.45%) and no other emergent adverse events.

Conclusions: The miCOR handheld non-cavitating lensectomy system demonstrated nuclear fragmentation and extraction in the absence of intraocular cavitation across all grades of nuclear densities.

PubMed Disclaimer

References

    1. Allen D, Vasavada A. Cataract and surgery for cataract. BMJ 2006;333:128–132
    1. Davis G. The evolution of cataract surgery. Mo Med 2016;113:58–62
    1. Liu YC, Wilkins M, Kim T, Malyugin B, Mehta JS. Cataracts. Lancet 2017;390:600–612
    1. Khanna R, Pujari S, Sangwan V. Cataract surgery in developing countries. Curr Opin Ophthalmol 2011;22:10–14
    1. Dewey S, Beiko G, Braga-Mele R, Nixon DR, Raviv T, Rosenthal K; ASCRS Cataract Clinical Committee; Instrumentation and IOLs Subcommittee. Microincisions in cataract surgery. J Cataract Refract Surg 2014;40:1549–1557

LinkOut - more resources