Accuracy of cup placement and pelvic motion in total hip arthroplasty in the lateral decubitus position using a new computed tomography-based navigation system with augmented reality technology
- PMID: 38554208
- DOI: 10.1007/s00402-024-05284-8
Accuracy of cup placement and pelvic motion in total hip arthroplasty in the lateral decubitus position using a new computed tomography-based navigation system with augmented reality technology
Abstract
Introduction: This study aimed to investigate the accuracy of cup position and assess the changes in pelvic tilt during primary total hip arthroplasty (THA) in the lateral decubitus position using a new computed tomography (CT)-based navigation system with augmented reality (AR) technology.
Materials and methods: There were 37 cementless THAs performed using a CT-based navigation system with AR technology in the lateral decubitus position and 63 cementless THAs performed using manual implant techniques in the lateral decubitus position in this retrospective study. Postoperative cup radiographic inclination and anteversion were measured using postoperative CT, and the proportion of hips within Lewinnek's safe zone was analyzed and compared between the two groups. The mean absolute values of navigation error were assessed. Intraoperative pelvic tilt angles were also recorded using navigation system.
Results: The percentage of cups inside Lewinnek's safe zone was 100% in the navigation group and 35% in the control group (p < 0.001). The mean absolute values of navigation error in inclination and anteversion were 2.9° ± 2.1° and 3.3° ± 2.4°, respectively. The mean abduction angle of the pelvis was 5.1° ± 4.8° after placing the patients in the lateral decubitus position and 4.1° ± 6.0° after cup placement. The mean posterior tilt angle was 6.8° ± 5.1° after placing the patients in the lateral decubitus position and 9.3° ± 5.9° after cup placement. The mean internal rotation angle was 14.8° ± 7.4° after cup placement. There were no correlations between the navigation error in inclination or anteversion and the absolute values of changes of the pelvic tilt angle at any phase.
Conclusions: Although progressive pelvic motion occurred in THA in the lateral decubitus position, especially during cup placement, the CT-based navigation system with AR technology improved cup placement accuracy.
Keywords: Augmented reality technology; Computed tomography-based navigation; Lateral decubitus position; Pelvic motion; Total hip arthroplasty.
© 2024. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature.
Similar articles
-
A novel imageless accelerometer-based navigation system improves acetabular cup placement accuracy during total hip arthroplasty in the lateral decubitus position.Arch Orthop Trauma Surg. 2024 Jun;144(6):2865-2872. doi: 10.1007/s00402-024-05376-5. Epub 2024 May 16. Arch Orthop Trauma Surg. 2024. PMID: 38753013
-
Accelerometer-based portable navigation system shows no superior accuracy over pelvic alignment guide for acetabular cup placement in total hip arthroplasty in lateral decubitus position.Arch Orthop Trauma Surg. 2024 Nov;144(11):4849-4856. doi: 10.1007/s00402-024-05571-4. Epub 2024 Sep 30. Arch Orthop Trauma Surg. 2024. PMID: 39347968
-
Registration in the supine position improve the accuracy of cup placement in total hip arthroplasty using a portable navigation system.Sci Rep. 2023 Nov 18;13(1):20222. doi: 10.1038/s41598-023-47674-9. Sci Rep. 2023. PMID: 37980455 Free PMC article.
-
Robotic arm-assisted total hip arthroplasty via a minimally invasive anterolateral approach in the supine position improves the precision of cup placement in patients with developmental dysplasia of the hip.J Orthop Sci. 2024 Mar;29(2):559-565. doi: 10.1016/j.jos.2023.01.012. Epub 2023 Feb 15. J Orthop Sci. 2024. PMID: 36801090 Review.
-
A high body mass index tilts the pelvis caudally in the lateral decubitus position for total hip arthroplasty.Hip Int. 2023 May;33(3):371-376. doi: 10.1177/11207000221084627. Epub 2022 Mar 20. Hip Int. 2023. PMID: 35311359 Review.
Cited by
-
Implant position accuracy using dynamic computer-assisted implant surgery (CAIS) combined with augmented reality: A randomized controlled clinical trial.J Dent Sci. 2024 Dec;19(Suppl 1):S44-S50. doi: 10.1016/j.jds.2024.09.004. Epub 2024 Sep 26. J Dent Sci. 2024. PMID: 39807435 Free PMC article.
-
Accuracy of Acetabular Component Positioning Using a Mixed Reality-Guided Navigation System During Total Hip Arthroplasty.JB JS Open Access. 2025 Mar 21;10(1):e24.00144. doi: 10.2106/JBJS.OA.24.00144. eCollection 2025 Jan-Mar. JB JS Open Access. 2025. PMID: 40124282 Free PMC article.
-
Accuracy of leg length changes in total hip arthroplasty using a computed tomography-based augmented reality navigation system.Arch Orthop Trauma Surg. 2024 Dec 12;145(1):17. doi: 10.1007/s00402-024-05705-8. Arch Orthop Trauma Surg. 2024. PMID: 39666089
References
-
- D’Lima DD, Urquhart AG, Buehler KO, Walker RH, Colwell CW Jr (2000) The effect of the orientation of the acetabular and femoral components on the range of motion of the hip at different head-neck ratios. J Bone Joint Surg Am 82:315–321. https://doi.10.2106/00004623-200003000-00003
-
- Jolles BM, Zangger P, Leyvraz PF (2002) Factors predisposing to dislocation after primary total hip arthroplasty: a multivariate analysis. J Arthroplasty 17:282–288. https://doi.10.1054/arth.2002.30286
-
- Kennedy JG, Rogers WB, Soffe KE, Sullivan RJ, Griffen DG, Sheehan LJ (1998) Effect of acetabular component orientation on recurrent dislocation, pelvic osteolysis, polyethylene wear, and component migration. J Arthroplasty 13:530–534. https://doi.10.1016/s0883-5403(98)90052-3
-
- Callanan MC, Jarrett B, Bragdon CR, Zurakowski D, Rubash HE, Freiberg AA, Malchau H (2011) The John Charnley Award: risk factors for cup malpositioning: quality improvement through a joint registry at a tertiary hospital. Clin Orthop Relat Res 469:319–329. https://doi.org/10.1007/s11999-010-1487-1
-
- Grammatopoulos G, Gofton W, Cochran M, Dobransky J, Carli A, Abdelbary H, Gill HS, Beaulé PE (2018) Pelvic positioning in the supine position leads to more consistent orientation of the acetabular component after total hip arthroplasty. Bone Joint J 100:1280–1288. https://doi.10.1302/0301-620X.100B10.BJJ-2018-0134.R1
MeSH terms
LinkOut - more resources
Full Text Sources
Medical
Research Materials