Comparative diagnostic accuracy of transcranial Doppler and contrast-enhanced transthoracic echocardiography for the diagnosis of patent foramen ovale and atrial septal defect
- PMID: 38563605
- DOI: 10.23736/S0031-0808.24.05123-1
Comparative diagnostic accuracy of transcranial Doppler and contrast-enhanced transthoracic echocardiography for the diagnosis of patent foramen ovale and atrial septal defect
Abstract
Background: Recognition of right-to-left shunt is crucial in the work-up of patients with suspected patent foramen ovale (PFO) or atrial septal defect (ASD). While transesophageal echocardiography (TEE) remains the gold standard diagnostic tool for the anatomic assessment of PFO/ASD, transcranial Doppler (TCD) and contrast-enhanced transthoracic echocardiogram (CE-TTE) hold the promise of providing minimally invasive yet accurate clinical details. Their comparative accuracy remains however debated.
Methods: We conducted a retrospective observational study leveraging our extensive institutional experience with systematic TCD and CE-TTE in patients with suspected PFO/ASD. Several measures of diagnostic test accuracy were computed, with point estimates and 95% confidence intervals, when applicable.
Results: A total of 1358 patients were included, with age 48±14 years and 772 (58%) women. Tests were performed for diagnostic purposes in 797 (58.6%) and during follow-up in 740 (54.5%). A PFO was eventually diagnosed in 1038 (77.9%) patients, and an ASD in 60 (4.5%). Agreement between TCD and CE-TTE occurred in 1309 (85.2%) cases, with TCD yielding worse findings than CE-TTE in 91 (5.9%) patients, and vice versa in 137 (8.9%), yielding a Cohen kappa of 78.6% (95% CI: 76.3-81.1%) and a highly significant P value at McNemar test (P<0.001). After dichotomization, and using TCD as benchmark, CE-TTE yielded sensitivity 96.9%, specificity 95.1%, area under the curve 92.1%, and P=0.249. Similar findings were obtained when focusing only on diagnostic tests or follow-up ones (Cohen kappa respectively 74.0% [70.2-77.1%], P<0.001 and 80.3% [76.4-84.3%], P<0.001). Notably, Valsalva was necessary to disclose the presence of shunt during TCD in 487 (31.7%) patients and during CE-TTE in 482 (31.4%) cases. Finally, performance of TCD and CE-TTE in a subset of patients eventually undergoing TTE was quite similar.
Conclusions: The diagnostic accuracy of CE-TTE appears favorable, and this imaging test may identify patients who may be missed if only TCD is used to screen patients with suspected PFO/ASD. Accordingly, CE-TTE is recommended as an adjunct diagnostic modality for all patients with a high pre-test probability of PFO/ASD and right-to-left shunt.
Similar articles
-
Transthoracic second harmonic two- and three-dimensional echocardiography for detection of patent foramen ovale.Eur J Echocardiogr. 2010 Jan;11(1):57-63. doi: 10.1093/ejechocard/jep165. Epub 2009 Nov 12. Eur J Echocardiogr. 2010. PMID: 19910318
-
Transcranial Doppler ultrasonography as a screening technique for detection of a patent foramen ovale before surgery in the sitting position.Anesthesiology. 2000 Oct;93(4):971-5. doi: 10.1097/00000542-200010000-00016. Anesthesiology. 2000. PMID: 11020748 Clinical Trial.
-
Diagnosis and quantification of patent foramen ovale. Which is the reference technique? Simultaneous study with transcranial Doppler, transthoracic and transesophageal echocardiography.Rev Esp Cardiol. 2011 Feb;64(2):133-9. doi: 10.1016/j.recesp.2010.10.009. Rev Esp Cardiol. 2011. PMID: 21277667 English, Spanish.
-
Transcranial Doppler versus transthoracic echocardiography for the detection of patent foramen ovale in patients with cryptogenic cerebral ischemia: A systematic review and diagnostic test accuracy meta-analysis.Ann Neurol. 2016 Apr;79(4):625-35. doi: 10.1002/ana.24609. Epub 2016 Mar 11. Ann Neurol. 2016. PMID: 26833864
-
The diagnostic value of contrast-enhanced transcranial Doppler and contrast-enhanced transthoracic echocardiography for right to left shunt in patent foramen ovale: a systematic review and meta-analysis.Front Neurol. 2024 Aug 2;15:1447964. doi: 10.3389/fneur.2024.1447964. eCollection 2024. Front Neurol. 2024. PMID: 39157064 Free PMC article.
Cited by
-
Comparison of the Application Value of Transthoracic Echocardiography in Diagnosing Patent Foramen Ovale Under Different States of Stimulation: A Retrospective Study.Clin Cardiol. 2024 Aug;47(8):e24319. doi: 10.1002/clc.24319. Clin Cardiol. 2024. PMID: 39109504 Free PMC article.
-
Genetic variation in patent foramen ovale: a case-control genome-wide association study.Front Genet. 2025 Jan 13;15:1523304. doi: 10.3389/fgene.2024.1523304. eCollection 2024. Front Genet. 2025. PMID: 39872005 Free PMC article.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources