Biomechanical Performance of Transtibial Pull-Out Posterior Horn Medial Meniscus Root Repair Is Improved With Knotless Adjustable Suture Anchor-Based Fixation
- PMID: 38584990
- PMCID: PMC10998488
- DOI: 10.1177/23259671241239575
Biomechanical Performance of Transtibial Pull-Out Posterior Horn Medial Meniscus Root Repair Is Improved With Knotless Adjustable Suture Anchor-Based Fixation
Abstract
Background: While posterior medial meniscus root (PMMR) techniques have evolved, there remains a need to both optimize repair strength and improve resistance to cyclic loading.
Hypothesis: Adjustable tensioning would lead to higher initial repair strength and reduce displacement with cyclic loading compared with previously described transtibial pull-out repair (TPOR) fixation techniques.
Study design: Controlled laboratory study.
Methods: A total of 56 porcine medial menisci were used. Eight intact specimens served as a control for the native meniscus. For the others, PMMR tears were created and repaired with 6 different TPOR techniques (8 in each group). Fixed PMMR repairs were executed using 4 different suture techniques (two No. 2 cinch sutures, two cinch tapes, two No. 2 simple sutures, and two No. 2 sutures in a Mason-Allen configuration) all tied over a cortical button. Adjustable PMMR repairs using Mason-Allen sutures were fixed with an adjustable soft tissue anchor fixation tensioned at either 80 N or 120 N. The initial force, stiffness, and relief displacement of the repairs were measured after fixation. Repair constructs were then cyclically loaded, with cyclic displacement and stiffness measured after 1000 cycles. Finally, the specimens were pulled to failure.
Results: The PMMR repaired with the 2 cinch sutures fixed technique afforded the lowest (P < .001) initial repair load, stiffness, and relief displacement. The adjustable PMMR repairs achieved a higher initial repair load (P < .001) and relief displacement (P < .001) than all fixed repairs. The 2 cinch sutures fixed technique showed an overall higher cyclic displacement (P < .028) and was completely loose compared with the native meniscus functional zone. Repairs with adjustable intratunnel fixation showed displacement with cyclic loading similar to the native meniscus. With cyclic loading, the Mason-Allen adjustable repair with 120 N of tension showed less displacement (P < .016) than all fixed repairs and a stiffness comparable to the fixed Mason-Allen repair. The fixed Mason-Allen technique demonstrated a higher ultimate load (P < .007) than the adjustable Mason-Allen techniques. All repairs were less stiff, with lower ultimate failure loads, than the native meniscus root attachment (P < .0001).
Conclusion: Adjustable TPOR led to considerably higher initial repair load and relief displacement than other conventional fixed repairs and restricted cyclic displacement to match the native meniscus function. However, the ultimate failure load of the adjustable devices was lower than that of a Mason-Allen construct tied over a cortical button. All repair techniques had a significantly lower load to failure than the native meniscus root.
Clinical relevance: Knotless adjustable PMMR repair based on soft anchor fixation results in higher tissue compression and less displacement, but the overall clinical significance on healing rates remains unclear.
Keywords: adjustable pull-out repair; biomechanics of ligament; knee; meniscus; meniscus root; root tear.
© The Author(s) 2024.
Conflict of interest statement
One or more of the authors has declared the following potential conflicts of interest: S.B. is an employee of Arthrex. C.A.W. is an employee of Arthrex. A.J.K. is a consultant for Arthrex and has received nonconsulting fees from Arthrex; consulting fees from Arthrex, Joint Restoration Foundation, and Responsive Arthroscopy LLC; royalties or license from Arthrex; grant from DJO LLC; and honoraria from Joint Restoration Foundation and Musculoskeletal Transplant Foundation. P.A.S. is a consultant for Arthrex and has received royalties or license from Arthrex; nonconsulting fees from Kairos Surgical; education payments from United Orthopedics LLC and Elite Orthopedics LLC; hospitality payments from Medical Device Business Services; and compensation for serving as faculty or as a speaker for an accredited or certified continuing education program from Medical Device Business Services. M.J.F. is a consultant for Arthrex. R.F.L. has received royalties or license from Arthrex and Smith+Nephew; education payments from Foundation Medical LLC; nonconsulting fees from Smith+Nephew; compensation for serving as faculty or as a speaker for a nonaccredited and noncertified continuing education program from Linvatec; and consulting fees from Smith+Nephew and Arthrex. AOSSM checks author disclosures against the Open Payments Database (OPD). AOSSM has not conducted an independent investigation on the OPD and disclaims any liability or responsibility relating thereto. Ethical approval was not sought for the present study.