Review
. 2025 Dec;32(6):1889-1905.
doi: 10.1177/15266028241245911. Epub 2024 Apr 16.

Effectiveness and Safety of Endoanchors in Abdominal and Thoracic Endovascular Aneurysm Repair: A Systematic Review and Meta-analysis

Affiliations
Review

Effectiveness and Safety of Endoanchors in Abdominal and Thoracic Endovascular Aneurysm Repair: A Systematic Review and Meta-analysis

Esther Wai Chi Tang et al. J Endovasc Ther. 2025 Dec.

Abstract

Purpose: This systematic review aimed to summarize the effectiveness and safety of endoanchor, a stabilizing device for the proximal endograft designed to prevent endoleak and stent migration in endovascular aneurysm repair (EVAR) and thoracic endovascular aneurysm repair (TEVAR).

Materials and methods: A systematic review and meta-analysis was conducted per the Preferred Reporting Items for Systematic Reviews and Meta-Analysis (PRISMA) guideline. Literature up to May 31, 2023 was searched and independently screened from 4 databases. Data were pooled for meta-analysis. Primary outcomes included intraoperative and follow-up endoleak, stent migration, and reintervention rates; sac regression; and 30-day all-cause mortality.

Results: Sixteen EVAR (n=1145) and 6 TEVAR studies (n=163) using the Heli-Fx EndoAnchor system were included from 2225 retrieved records. For EVAR patients (mean follow-up=11.9 months), the endoleak, graft migration, and reintervention rates were 3.97% (95% confidence interval [CI]=0.36%-1.99%), 0.004% (95% CI=0.00%-0.76%), and 5.43% (95% CI=0.86%-12.54%), respectively. The endoleak rates for primary and revision EVAR were 0.16% (95% CI=0.00%-1.65%) and 3.60% (95% CI=0.14%-9.72%), respectively. Only 4 cases of 30-day mortality (n=4) were reported in the literature. For TEVAR patients, the endoleak, stent migration, and reintervention rates were 7.4% (95% CI=0.03%-0.13%), 0.2% (95% CI=0.00%-0.06%), and 17.1% (95% CI=0.01%-0.45%), respectively. The 30-day mortality was 0.9% (95% CI=0%-0.12%).

Conclusions: Endoanchor fixation in EVAR and TEVAR is effective and safe in preventing and treating endoleak and stent migration. The mortality is minimal in EVAR but higher in TEVAR.Clinical ImpactEndoleak, graft migration, and reintervention in EVAR and TEVAR with endoanchor use were rare. Mortality in EVAR was low. The adjunctive deployment of endoanchors is an effective and safe means to prevent and treat endoleak and stent migration in EVAR and TEVAR. Yet, long-term efficacy and safety data and randomized controlled trials would be required to definitively recommend endoanchor use in routine clinical practice.

Keywords: endoanchor; endoleak; endovascular aneurysm repair; mortality; reintervention; stent migration; thoracic endovascular aneurysm repair.