Multicenter Study
. 2024 Jun:298:316-324.
doi: 10.1016/j.jss.2024.03.033. Epub 2024 Apr 18.

Clinical Safety and Performance of GATT-Patch for Hemostasis in Minimal to Moderate Bleeding During Open Liver Surgery

Affiliations
Free article
Multicenter Study

Clinical Safety and Performance of GATT-Patch for Hemostasis in Minimal to Moderate Bleeding During Open Liver Surgery

Johannes H W de Wilt et al. J Surg Res. 2024 Jun.
Free article

Abstract

Introduction: Intraoperative blood loss and postoperative hemorrhage affect outcomes after liver resection. GATT-Patch is a new flexible, pliable hemostatic sealant patch comprising fibrous gelatin carrier impregnated with N-hydroxy-succinimide polyoxazoline. We evaluated safety and performance of the GATT-Patch for hemostasis at the liver resection plane.

Methods: Adult patients undergoing elective open liver surgery were recruited in three centers. GATT-Patch was used for minimal to moderate bleeding at the liver resection plane. The primary endpoint was hemostasis of the first-treated bleeding site at 3 min versus a prespecified performance goal of 65.4%.

Results: Two trial stages were performed: I (n = 8) for initial safety and II (n = 39) as the primary outcome cohort. GATT-Patch was applied in 47 patients on 63 bleeding sites. Median age was 60.0 (range 25-80) years and 70% were male. Most (66%) surgeries were for colorectal cancer metastases. The primary endpoint was met in 38 out of 39 patients (97.4%; 95% confidence interval: 84.6%-99.9%) versus 65.4% (P < 0.001). Of all the 63 bleeding sites, hemostasis was 82.7% at 30, 93.7% at 60, and 96.8% at 180 s. No reoperations for rebleeding or device-related issues occurred.

Conclusions: When compared to a performance goal derived from state-of-the-art hemostatic agents, GATT-Patch for the treatment of minimal to moderate bleeding during liver surgery successfully and quickly achieved hemostasis with acceptable safety outcomes. (ClinicalTrials.gov Identifier: NCT04819945).

Keywords: Bleeding; GATT-Patch; Hemostasis; Hepatectomy; Liver surgery.