Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2024 Jul;25(7):e14370.
doi: 10.1002/acm2.14370. Epub 2024 Apr 25.

Electron streaming dose measurements and calculations on a 1.5 T MR-Linac

Affiliations

Electron streaming dose measurements and calculations on a 1.5 T MR-Linac

Elizabeth Patterson et al. J Appl Clin Med Phys. 2024 Jul.

Abstract

Purpose: To evaluate the accuracy of different dosimeters and the treatment planning system (TPS) for assessing the skin dose due to the electron streaming effect (ESE) on a 1.5 T magnetic resonance (MR)-linac.

Method: Skin dose due to the ESE on an MR-linac (Unity, Elekta) was investigated using a solid water phantom rotated 45° in the x-y plane (IEC61217) and centered at the isocenter. The phantom was irradiated with 1 × 1, 3 × 3, 5 × 5, 10 × 10, and 22 × 22 cm2 fields, gantry at 90°. Out-of-field doses (OFDs) deposited by electron streams generated at the entry and exit surface of the angled phantom were measured on the surface of solid water slabs placed ±20.0 cm from the isocenter along the x-direction. A high-resolution MOSkin™ detector served as a benchmark due to its shallower depth of measurement that matches the International Commission on Radiological Protection (ICRP) recommended depth for skin dose assessment (0.07 mm). MOSkin™ doses were compared to EBT3 film, OSLDs, a diamond detector, and the TPS where the experimental setup was modeled using two separate calculation parameters settings: a 0.1 cm dose grid with 0.2% statistical uncertainty (0.1 cm, 0.2%) and a 0.2 cm dose grid with 3.0% statistical uncertainty (0.2 cm, 3.0%).

Results: OSLD, film, the 0.1 cm, 0.2%, and 0.2 cm, 3.0% TPS ESE doses, underestimated skin doses measured by the MOSkin™ by as much as -75.3%, -7.0%, -24.7%, and -41.9%, respectively. Film results were most similar to MOSkin™ skin dose measurements.

Conclusions: These results show that electron streams can deposit significant doses outside the primary field and that dosimeter choice and TPS calculation settings greatly influence the reported readings. Due to the steep dose gradient of the ESE, EBT3 film remains the choice for accurate skin dose assessment in this challenging environment.

Keywords: Elekta Unity; MRgRT; MR‐linac; Monaco Unity; electron streaming effect; out‐of‐field dose; skin dose.

PubMed Disclaimer

Conflict of interest statement

D.Cutajar declares consulting with Electrogenics Laboratories Ltd which is commercializing the MOSkin™ detector. Elizabeth Patterson, Marcus Powers, Peter E. Metcalfe, Bradley M. Oborn, and John A. Baines have no relevant conflicts of interest to disclose.

Figures

FIGURE 1
FIGURE 1
Schematic of the setup used to generate backscattered and ejected ESE.
FIGURE 2
FIGURE 2
(a) Photograph of the experimental setup with the MOSkin™ positioned in the holder placed 20.0 cm from the isocenter to measure ejected ESE. (b) Photograph of the microDiamond positioned using a customized phantom to measure ejected ESE.
FIGURE 3
FIGURE 3
(a) A coronal slice of the calculated dose distribution for a G90, 1000 MU 10 × 10 cm2 beam irradiating the angled solid water. (b) Ejected ESE transverse dose profiles in the x‐direction for a 1 × 1 cm2 and 10 × 10 cm2 beam at the surface of the +y positioned solid water phantom. Similarly in (c) for backscattered ESE at the ‐y positioned block. The dashed vertical line in (b) and (c) indicates the position used for comparison between experimental and TPS computed datasets. All data is normalized to Dmax of a 10 × 10 cm2 field.
FIGURE 4
FIGURE 4
Measured and TPS predicted backscattered ESE doses for (a) 1 × 1 cm2, (b) 3 × 3 cm2, (c) 5 × 5 cm2, (d) 10 × 10 cm2, and (e) 22 × 22 cm2 fields. All data is normalized to Dmax of a 10 × 10 cm2 field. MOSkin™ (×), OSLD (○), film (□), microDiamond (+), TPS (0.1 cm, 0.2%) (△), and TPS (0.2 cm, 3.0%) (◇) point doses are denoted by their respective marker whilst the dashed line represents curve‐fit MOSkin™ doses. For each dosimeter measurement, the estimated error is given by an error bar.
FIGURE 5
FIGURE 5
Measured and TPS predicted ejected ESE doses for (a) 1 × 1 cm2, (b) 3 × 3 cm2, (c) 5 × 5 cm2, (d) 10 × 10 cm2, and (e) 22 × 22 cm2 fields. All data is normalized to Dmax of a 10 × 10 cm2 field. MOSkin™ (×), OSLD (○), film (□), microDiamond (+), TPS (0.1 cm, 0.2%) (△), and TPS (0.2 cm, 3.0%) (◇) point doses are denoted by their respective marker while the dashed line represents curve‐fit MOSkin™ doses. For each dosimeter measurement, the estimated error is given by an error bar.
FIGURE 6
FIGURE 6
Box plot of the relative difference, ΔD, between the 0.07 mm WED MOSkin™ recorded dose values and OSLDs, film, and planning system values for (a) backscattered ESE, and (b) ejected ESE doses across all five field sizes investigated. For each box plot the red line indicates the median value, the box indicates the 25th–75th percentiles, the bar indicates extreme data points, and the red crosses indicate outliers.

References

    1. Keall PJ, Glide‐Hurst CK, Cao M, et al. ICRU REPORT 97: MRI‐guided radiation therapy using MRI‐linear accelerators. J ICRU. 2022;22(1):1‐100. 10.1177/14736691221141950 - DOI
    1. Raaymakers BW, Jürgenliemk‐Schulz IM, Bol GH, et al. First patients treated with a 1.5 T MRI‐linac: clinical proof of concept of a high‐precision, high‐field MRI guided radiotherapy treatment. Phys Med Biol. 2017;62(23):L41‐L50. 10.1088/1361-6560/AA9517 - DOI - PubMed
    1. Mutic S, Dempsey JF. The ViewRay system: magnetic resonance‐guided and controlled radiotherapy. Semin Radiat Oncol. 2014;24(3):196‐199. 10.1016/J.SEMRADONC.2014.02.008 - DOI - PubMed
    1. Yang B, Tang KK, Huang CY, et al. Out‐of‐field dose and its constituent components for a 1.5 T MR‐Linac. Phys Med Biol. 2021;66(22):225012. 10.1088/1361-6560/AC3346 - DOI - PubMed
    1. Hackett SL, Van Asselen B, Wolthaus JWH, et al. Spiraling contaminant electrons increase doses to surfaces outside the photon beam of an MRI‐linac with a perpendicular magnetic field. Phys Med Biol. 2018;63(9):095001. 10.1088/1361-6560/aaba8f - DOI - PubMed

MeSH terms

LinkOut - more resources