. 2024 Oct;27(5):758-766.
doi: 10.1111/ocr.12801. Epub 2024 Apr 29.

Evaluation of Long-term colour resistance of composite materials used in aligner treatments: A comparative study

Affiliations
Comparative Study

Evaluation of Long-term colour resistance of composite materials used in aligner treatments: A comparative study

Hilal Yildiz et al. Orthod Craniofac Res. 2024 Oct.

Abstract

Aim: The aim of this study is to evaluate the discolouration resistance of attachments made of composites having different physical properties, filler ratios and viscosity through the simulated environment and conditions.

Materials and methods: Seventy-two extracted human teeth were aligned to simulate upper and lower dental arches. The attachments were prepared according to the guides determined by ClearCorrect. Three different composite groups were used for attachments preparation: Aligner Connect, G-Aenial Universal Injectable, and Tetric Prime and Tetric Evoflow combination. Dental models and aligners were kept in artificial saliva throughout the experiment of 96 weeks. The aligners were also subjected to 2308 thermal cycles and insertion and removing procedures 840 times for each period of 12 weeks. Afterwards, the models were kept in coffee and red wine solutions. Colour measurements were made with Vita Easy Shade 4.0 at 12th, 24th, 48th and 96th simulated weeks.

Results: No significant difference in discolouration was observed during thermal cycle application of 12th and 24th week experimental periods. In the later periods, discolouration was observed in the composite materials, and G-Aenial Universal Injectable was less coloured than the other samples. The composite samples were more coloured in proportion to their residence time. The G-Aenial Universal Injectable was more affected by wine solution, and Tetric Prime and Tetric Evoflow combination were more affected by coffee solution.

Conclusions: Viscosity, resin matrix content and nature of inorganic filler particles of the composite materials significantly affect the colour changing properties of them due to the long-term ageing and exposure to colouring agents.

Keywords: clear aligners; colour stability; composite attachments; discolouration; orthodontic appliance.

PubMed Disclaimer

References

REFERENCES

    1. D'Antò V, Muraglie S, Castellano B, et al. Influence of dental composite viscosity in attachment reproduction: an experimental in vitro study. Materials (Basel). 2019;12(23):4001.
    1. Morton J, Derakhshan M, Kaza S, Li C. Design of the Invisalign system performance. Semin Orthod. 2017;23:3‐11.
    1. Mantovani E, Castroflorio E, Rossini G, et al. Scanning electron microscopy analysis of aligner fitting on Anchorage attachments. J Orofac Orthop. 2019;80(2):79‐87.
    1. Labella R, Lambrechts P, Van Meerbeek B, Vanherle G. Polymerization shrinkage and elasticity of Flowable composites and filled adhesives. Dent Mater. 1999;15(2):128‐137.
    1. Gazzani F, Bellisario D, Quadrini F, et al. Comparison between different composite resins used for clear aligner attachments: an in‐vitro study. Front Mater. 2022;8:789143.

Publication types

LinkOut - more resources