Multicenter Study
. 2024 May 3;65(5):ezae122.
doi: 10.1093/ejcts/ezae122.

Examining the typical hemodynamic performance of nearly 3000 modern surgical aortic bioprostheses

Affiliations
Multicenter Study

Examining the typical hemodynamic performance of nearly 3000 modern surgical aortic bioprostheses

Robert J M Klautz et al. Eur J Cardiothorac Surg. .

Abstract

Objectives: The objective of this analysis was to assess the normal haemodynamic performance of contemporary surgical aortic valves at 1 year postimplant in patients undergoing surgical aortic valve replacement for significant valvular dysfunction. By pooling data from 4 multicentre studies, this study will contribute to a better understanding of the effectiveness of surgical aortic valve replacement procedures, aiding clinicians and researchers in making informed decisions regarding valve selection and patient management.

Methods: Echocardiograms were assessed by a single core laboratory. Effective orifice area, dimensionless velocity index, mean aortic gradient, peak aortic velocity and stroke volume were evaluated.

Results: The cohort included 2958 patients. Baseline age in the studies ranged from 70.1 ± 9.0 to 83.3 ± 6.4 years, and Society of Thoracic Surgeons risk of mortality was 1.9 ± 0.7 to 7.5 ± 3.4%. Twenty patients who had received a valve model implanted in fewer than 10 cases were excluded. Ten valve models (all tissue valves; n = 2938 patients) were analysed. At 1 year, population mean effective orifice area ranged from 1.46 ± 0.34 to 2.12 ± 0.59 cm2, and dimensionless velocity index, from 0.39 ± 0.07 to 0.56 ± 0.15. The mean gradient ranged from 8.6 ± 3.4 to 16.1 ± 6.2 mmHg with peak aortic velocity of 1.96 ± 0.39 to 2.65 ± 0.47 m/s. Stroke volume was 75.3 ± 19.6 to 89.8 ± 24.3 ml.

Conclusions: This pooled cohort is the largest to date of contemporary surgical aortic valves with echocardiograms analysed by a single core lab. Overall haemodynamic performance at 1 year ranged from good to excellent. These data can serve as a benchmark for other studies and may be useful to evaluate the performance of bioprosthetic surgical valves over time.

Clinical trial registration number: NCT02088554, NCT02701283, NCT01586910 and NCT01531374.

Keywords: Aortic valve replacement; Surgical aortic bioprostheses; Valve haemodynamic performance.