[The Myth about the Laryngeal Mask]
- PMID: 38759687
- DOI: 10.1055/a-2199-4692
[The Myth about the Laryngeal Mask]
Abstract
Peak pressures ≥ 20 mbar are not a contraindication for laryngeal masks. The oropharyngeal leak pressure of a laryngeal mask does not correspond to the pressure at which oesophagogastric air leakage occurs. Setting a peak pressure limit of 20 cm H2O on the respirator can lead to critical situations because the tidal volume may then remain too low. A good alternative is to use a pressure alarm limit. The use of laryngeal masks does not preclude the use of PEEP and/or relaxation.
Thieme. All rights reserved.