Neurovascular Imaging with Ultra-High-Resolution Photon-Counting CT: Preliminary Findings on Image-Quality Evaluation
- PMID: 38760079
- PMCID: PMC11448984
- DOI: 10.3174/ajnr.A8350
Neurovascular Imaging with Ultra-High-Resolution Photon-Counting CT: Preliminary Findings on Image-Quality Evaluation
Abstract
Background and purpose: The first-generation photon-counting detector CT was recently introduced into clinical practice and represents a promising innovation in high-resolution CT imaging. The purpose of this study was to assess the image quality of ultra-high-resolution photon-counting detector CT compared with energy-integrating detector CT and to explore different reconstruction kernel sharpness levels for the evaluation of intracranial aneurysms.
Materials and methods: Ten patients with intracranial saccular aneurysms who had previously undergone conventional energy-integrating detector CT were prospectively enrolled. CT angiograms were acquired on a clinical dual-source photon-counting detector CT in ultra-high-resolution mode and reconstructed with 4 vascular kernels (Bv36, Bv40, Bv44, Bv48). Quantitative and qualitative image-quality parameters of the intracranial arteries were evaluated. For the quantitative analysis (image noise, SNR, contrast-to-noise ratio), ROIs were manually placed at standard anatomic intracranial and extracranial locations by 1 author. In addition, vessel border sharpness was evaluated quantitatively. For the qualitative analysis, 3 blinded neuroradiologists rated photon-counting detector CT and energy-integrating detector CT image quality for the evaluation of the intracranial vessels (ie, the aneurysms and 9 standard vascular branching locations) on a 5-point Likert-type scale. Additionally, readers independently selected their preferred kernel among the 4 kernels evaluated on photon-counting detector CT.
Results: In terms of quantitative image quality, Bv48, the sharpest kernel, yielded increased image noise and decreased SNR and contrast-to-noise ratio parameters compared with Bv36, the smoothest kernel. Compared with energy-integrating detector CT, the Bv48 kernel offered better quantitative image quality for the evaluation of small intracranial vessels (P < .001). Image-quality ratings of the Bv48 were superior to those of the energy-integrating detector CT and not significantly different from ratings of the B44 reconstruction kernel. When comparing side by side all 4 photon-counting detector reconstruction kernels, readers selected the B48 kernel as the best to visualize the aneurysms in 80% of cases.
Conclusions: Ultra-high-resolution photon-counting detector CT provides improved image quality for neurovascular imaging. Although the less sharp kernels provided superior SNR and contrast-to-noise ratio, the sharpest kernels delivered the best subjective image quality on photon-counting detector CT for the evaluation of intracranial aneurysms.
© 2024 by American Journal of Neuroradiology.
Similar articles
-
Imaging of Cardiovascular Implantable Electronic Device Leads With Photon-Counting Detector Computed Tomography.Invest Radiol. 2025 Sep 1;60(9):616-625. doi: 10.1097/RLI.0000000000001177. Invest Radiol. 2025. PMID: 40138655
-
High-Resolution Head CTA: A Prospective Patient Study Comparing Image Quality of Photon-Counting Detector CT and Energy-Integrating Detector CT.AJNR Am J Neuroradiol. 2024 Oct 3;45(10):1441-1449. doi: 10.3174/ajnr.A8342. AJNR Am J Neuroradiol. 2024. PMID: 39237360
-
Advancing Neurovascular Imaging: Optimization of Reconstruction Kernel and Quantum Iterative Reconstruction for Ultra-high-resolution Photon-Counting Detector CT Angiography of the Head and Neck.Acad Radiol. 2025 Aug 1:S1076-6332(25)00703-2. doi: 10.1016/j.acra.2025.07.033. Online ahead of print. Acad Radiol. 2025. PMID: 40753021
-
Photon-counting CT versus energy-integrating detectors for cardiac imaging: a systematic review of evidence from in vivo human studies on image quality and radiation dose.BMC Med Imaging. 2025 Jul 25;25(1):295. doi: 10.1186/s12880-025-01825-8. BMC Med Imaging. 2025. PMID: 40713483 Free PMC article.
-
Adefovir dipivoxil and pegylated interferon alfa-2a for the treatment of chronic hepatitis B: a systematic review and economic evaluation.Health Technol Assess. 2006 Aug;10(28):iii-iv, xi-xiv, 1-183. doi: 10.3310/hta10280. Health Technol Assess. 2006. PMID: 16904047
Cited by
-
Photon-counting detector CTA to assess intracranial stents and flow diverters: an in vivo study with ultrahigh-resolution spectral reconstructions.Eur Radiol Exp. 2025 Jan 29;9(1):10. doi: 10.1186/s41747-025-00550-9. Eur Radiol Exp. 2025. PMID: 39881008 Free PMC article.
-
Optimizing Photon-Counting Detector CT for Imaging Intracranial Aneurysms.AJNR Am J Neuroradiol. 2024 Oct 3;45(10):1458-1460. doi: 10.3174/ajnr.A8400. AJNR Am J Neuroradiol. 2024. PMID: 39122465 No abstract available.
-
Ultra-high-resolution imaging of intracranial flow diverters with photon counting CT: A comparative phantom study with flat-panel CT.Sci Rep. 2025 Jul 21;15(1):26498. doi: 10.1038/s41598-025-12713-0. Sci Rep. 2025. PMID: 40691292 Free PMC article.
-
Visibility of Intracranial Perforating Arteries Using Ultra-High-Resolution Photon-Counting Detector Computed Tomography (CT) Angiography.Tomography. 2024 Nov 21;10(12):1867-1880. doi: 10.3390/tomography10120136. Tomography. 2024. PMID: 39728898 Free PMC article.
-
Photon-counting CT imaging of a patient with coiled and untreated intracranial saccular aneurysms.Neuroradiol J. 2025 Jan 29:19714009251313514. doi: 10.1177/19714009251313514. Online ahead of print. Neuroradiol J. 2025. PMID: 39881477 Free PMC article.
References
MeSH terms
LinkOut - more resources
Full Text Sources
Medical