Abstract
Background: Central venous access devices (CVADs) allow intravenous therapy, haemodynamic monitoring and blood sampling but many fail before therapy completion.
Objective: To quantify CVAD failure and complications; and identify risk factors.
Designs, settings, and participants: Secondary analysis of multicentre randomised controlled trial including patients aged ≥16 years with a non-tunnelled CVAD (NTCVAD), peripherally-inserted central catheter (PICC) or tunnelled CVAD (TCVAD). Primary outcome was incidence of all-cause CVAD failure (central line-associated bloodstream infection [CLABSI], occlusion, accidental dislodgement, catheter fracture, thrombosis, pain). Secondary outcomes were CLABSI, occlusion and dislodgement. Cox regression was used to report time-to-event associations.
Results: In 1892 CVADs, all-cause failure occurred in 10.2% of devices: 49 NTCVADs (6.1%); 100 PICCs (13.2%); 44 TCVADs (13.4%). Failure rates for CLABSI, occlusion and dislodgement were 5.3%, 1.8%, and 1.7%, respectively. Independent CLABSI predictors were blood product administration through PICCs (hazard ratio (HR) 2.62, 95% confidence interval (CI) 1.24-5.55); and in TCVADs, one or two lumens, compared with three to four (HR 3.36, 95%CI 1.68-6.71), intravenous chemotherapy (HR 2.96, 95%CI 1.31-6.68), and diabetes (HR 3.25, 95%CI 1.40-7.57). Independent factors protective for CLABSI include antimicrobial NTCVADs (HR 0.23, 95%CI 0.08-0.63) and lipids in TCVADs (HR 0.32, 95%CI 0.14-0.72). NTCVADs inserted at another hospital (HR 7.06, 95%CI 1.48-33.7) and baseline infection in patients with PICCs (HR 2.72, 95%CI 1.08-6.83) were predictors for dislodgement. No independent occlusion predictors were found. Modifiable risk factors were identified for CVAD failure, which occurred for 1-in-10 catheters. Strict infection prevention measures and improved CVAD securement could reduce CLABSI and dislodgement risk.
© 2024 The Author(s). Journal of Hospital Medicine published by Wiley Periodicals LLC on behalf of Society of Hospital Medicine.
References
REFERENCES
-
- Loveday HP, Wilson JA, Pratt RJ, et al. epic3: national evidence‐based guidelines for preventing healthcare‐associated infections in NHS hospitals in England. J Hosp Infect. 2014;86(suppl 1):S1‐S70.
-
- Moureau NL, Alexandrou E. Device selection. In: Moreau NL, ed., Vessel Health and Presearvation: The Right Approach for Vascular Access. Springer; 2019.
-
- Chopra V, O'Horo JC, Rogers MAM, Maki DG, Safdar N. The risk of bloodstream infection associated with peripherally inserted central catheters compared with central venous catheters in adults: a systematic review and meta‐analysis. Infect Control Hosp Epidemiol. 2013;34(9):908‐918.
-
- Chou EH, Mann S, Hsu TC, et al. Incidence, trends, and outcomes of infection sites among hospitalizations of sepsis: a nationwide study. PLoS One. 2020;15(1):e0227752.
-
- Ullman AJ, Marsh N, Mihala G, Cooke M, Rickard CM. Complications of central venous access devices: a systematic review. Pediatrics. 2015;136(5):e1331‐e1344.