Exploring morphometric transformations in temperature-sensitive NiTi instruments alongside real-time intracanal temperature dynamics
- PMID: 38804711
- DOI: 10.1111/iej.14100
Exploring morphometric transformations in temperature-sensitive NiTi instruments alongside real-time intracanal temperature dynamics
Abstract
Aim: This study was twofold: (i) it aimed to investigate the morphometric changes of three temperature-sensitive nickel-titanium (NiTi) instruments at different temperatures, and (ii) to conduct an in vivo real-time analysis of intracanal temperature changes.
Methods: Changes in the shape and length of XP-Endo Shaper, XP-Endo Finisher, and XP-Endo Finisher-R were evaluated in real time whilst heated in a temperature-controlled water bath from 22 to 45°C. Instruments were fixed to a laminated water-resistant 1 mm graph paper attached to a stone block. Instruments were imaged whilst subjected to increasing temperature using a digital camera attached to an operating microscope. From recorded videos, still frames were extracted at 10-s intervals and changes in the length and shape of each instrument were measured and changes were plotted against time. Moreover, the intracanal temperature of distal roots of lower molars was measured in vivo for patients attending the clinic for non-surgical root canal treatments. The temperature was measured using a K-type thermocouple probe inserted into the mid-root level after irrigating the canal with a solution set at room temperature (22°C) or heated to 45°C. The intraoral and intracanal temperatures were recorded using a video camera for 180 s at 5-s intervals to plot the change in the intraoral and intracanal temperature, after both irrigation solution temperatures, with time.
Results: The shape transformation of XP-Endo Shaper began at 31.5 ± 2.0°C and reached its optimal transformation at 35.1 ± 1.0°C. For the Finisher and Finisher-R, shape transformations began at 29.2 ± 1.9 and 26.9 ± 2.2°C reaching the optimal transformation at 33.9 ± 1.4 and 32.7 ± 1.7°C, respectively. The average decreases in lengths of XP-Endo Shaper, Finisher, and Finisher-R after full transformation were 0.43 ± 0.23, 1.07 ± 0.22, and 1.15 ± 0.22 mm, respectively. The intracanal temperature reached 32.9 ± 0.8 and 33.2 ± 1.0°C after 3 min of application of irrigation solutions set at 22 or 45°C, respectively.
Conclusion: The tested instruments exhibited diverse changes in their shapes and lengths at varying temperatures. Despite the temperature of the irrigation solution, the intracanal temperature consistently remained lower than the intracanal temperature once equilibrium was reached. This highlights the importance of considering the temperature of irrigation solution during in vitro testing of endodontic instruments.
Keywords: NiTi instruments; intracanal temperature; irrigation temperature; shape transformation.
© 2024 British Endodontic Society. Published by John Wiley & Sons Ltd.
Similar articles
-
XP-Endo Finisher behaviour and irrigant temperature in intracanal environment: Clinical measurements.Int Endod J. 2025 Jan;58(1):141-149. doi: 10.1111/iej.14146. Epub 2024 Sep 12. Int Endod J. 2025. PMID: 39264747
-
Cyclic Fatigue Resistance of XP-Endo Shaper, K3XF, and ProTaper Gold Nickel-titanium Instruments.J Endod. 2018 Jul;44(7):1164-1167. doi: 10.1016/j.joen.2018.04.005. Epub 2018 Jun 1. J Endod. 2018. PMID: 29866409
-
Characterisation of deformed or separated nickel-titanium retreatment instruments after clinical use - A multicentre experience: Defect profiles of clinically-used retreatment instruments.J Dent. 2022 Feb;117:103939. doi: 10.1016/j.jdent.2021.103939. Epub 2021 Dec 23. J Dent. 2022. PMID: 34953972
-
Bacteria and Hard Tissue Debris Extrusion and Intracanal Bacterial Reduction Promoted by XP-endo Shaper and Reciproc Instruments.J Endod. 2018 Jul;44(7):1173-1178. doi: 10.1016/j.joen.2018.04.007. Epub 2018 Jun 1. J Endod. 2018. PMID: 29861066
-
Effectiveness of XP-Endo Finisher and passive ultrasonic irrigation on intracanal medicament removal from root canals: a systematic review and meta-analysis.BMC Oral Health. 2021 Jun 9;21(1):294. doi: 10.1186/s12903-021-01644-7. BMC Oral Health. 2021. PMID: 34107959 Free PMC article.
Cited by
-
The impact of blade geometry (nonflat, flat or hybrid) and metallurgical composition on the mechanical performance of NiTi endodontic instruments: A multimethod laboratory study.Int Endod J. 2025 Aug;58(8):1243-1256. doi: 10.1111/iej.14249. Epub 2025 May 3. Int Endod J. 2025. PMID: 40317569 Free PMC article.
-
Non-surgical endodontics: contemporary biomechanical preparation of the root canal system.Br Dent J. 2025 Apr;238(7):478-486. doi: 10.1038/s41415-025-8599-1. Epub 2025 Apr 11. Br Dent J. 2025. PMID: 40217030 Free PMC article. Review.
-
Assessing the fatigue resistance of NiTi instruments: A DSC-based approach to understanding temperature effects.Int Endod J. 2025 Aug;58(8):1267-1276. doi: 10.1111/iej.14257. Epub 2025 May 23. Int Endod J. 2025. PMID: 40407736 Free PMC article.
-
Characteristics and Effectiveness of XP-Endo Files and Systems: A Narrative Review.Int J Dent. 2024 Dec 17;2024:9412427. doi: 10.1155/ijod/9412427. eCollection 2024. Int J Dent. 2024. PMID: 39720541 Free PMC article. Review.
-
Influence of Continuous Rotation and Optimal Torque Reverse Kinematics on the Cyclic Fatigue Strength of Endodontic NiTi Clockwise Cutting Rotary Instruments.Dent J (Basel). 2024 Sep 30;12(10):317. doi: 10.3390/dj12100317. Dent J (Basel). 2024. PMID: 39452445 Free PMC article.
References
REFERENCES
-
- Boutsioukis, C. & Arias‐Moliz, M.T. (2022) Present status and future directions – irrigants and irrigation methods. International Endodontic Journal, 55(Suppl 3), 588–612. Available from: https://doi.org/10.1111/iej.13739
-
- Cunningham, W.T. & Balekjian, A.Y. (1980) Effect of temperature on collagen‐dissolving ability of sodium hypochlorite endodontic irrigant. Oral Surgery, Oral Medicine, and Oral Pathology, 49(2), 175–177.
-
- de Hemptinne, F., Slaus, G., Vandendael, M., Jacquet, W., De Moor, R.J. & Bottenberg, P. (2015) In vivo intracanal temperature evolution during endodontic treatment after the injection of room temperature or preheated sodium hypochlorite. Journal of Endodontics, 41(7), 1112–1115.
-
- de Vasconcelos, R.A., Murphy, S., Carvalho, C.A., Govindjee, R.G., Govindjee, S. & Peters, O.A. (2016) Evidence for reduced fatigue resistance of contemporary rotary instruments exposed to body temperature. Journal of Endodontics, 42(5), 782–787.
-
- De‐Deus, G., Belladonna, F.G., Silva, E.J., Marins, J.R., Souza, E.M., Perez, R. et al. (2015) Micro‐CT evaluation of non‐instrumented canal areas with different enlargements performed by NiTi systems. Brazilian Dental Journal, 26, 624–629.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources