Two-Dimensional Versus Three-Dimensional Preoperative Planning in Total Hip Arthroplasty
- PMID: 38810812
- DOI: 10.1016/j.arth.2024.05.054
Two-Dimensional Versus Three-Dimensional Preoperative Planning in Total Hip Arthroplasty
Abstract
Background: Preoperative planning in total hip arthroplasty (THA) involves utilizing radiographs or advanced imaging modalities, including computerized tomography scans, for precise prediction of implant sizing and positioning. This study aimed to compare 3-dimensional (3D) versus 2-dimensional (2D) preoperative planning in primary THA with respect to key surgical metrics, including restoration of the horizontal and vertical center of rotation (COR), combined offset, and leg length.
Methods: This study included 60 patients undergoing primary THA for symptomatic hip osteoarthritis (OA), randomly allocated to either robotic arm-assisted or conventional THA. Digital 2D templating and 3D planning using the robotic software were performed for all patients. All measurements to evaluate the accuracy of templating methods were conducted on the preoperative computerized tomography scanogram, using the contralateral hip as a reference. Sensitivity analyses explored differences between 2D and 3D planning in patients who had superolateral or medial OA patterns.
Results: Compared to 2D templating, 3D templating was associated with less medialization of the horizontal COR (-1.2 versus -0.2 mm, P = .002) and more accurate restoration of the vertical COR (1.63 versus 0.3 mm, P < .001) with respect to the contralateral side. Furthermore, 3D templating was superior for planned restoration of leg length (+0.23 versus -0.74 mm, P = .019). Sensitivity analyses demonstrated that in patients who had medial OA, 3D planning resulted in less medialization of horizontal COR and less offset reduction. Conversely, in patients who had superolateral OA, there was less lateralization of horizontal COR and less offset increase using 3D planning. Additionally, 3D planning showed superior reproducibility for stem, acetabular cup sizes, and neck angle, while 2D planning often led to smaller stem and cup sizes.
Conclusions: Our findings indicated higher accuracy in the planned restoration of native joint mechanics using 3D planning. Additionally, this study highlights distinct variances between the 2 planning methods across different OA pattern subtypes, offering valuable insights for clinicians employing 2D planning.
Keywords: CT-based planning; center of rotation; offset; randomized controlled trial; robotic arm assistance; total hip arthroplasty.
Copyright © 2024 Elsevier Inc. All rights reserved.
Similar articles
-
Accuracy of the preoperative planning for cementless total hip arthroplasty. A randomised comparison between three-dimensional computerised planning and conventional templating.Orthop Traumatol Surg Res. 2012 Apr;98(2):151-8. doi: 10.1016/j.otsr.2011.09.023. Epub 2012 Feb 29. Orthop Traumatol Surg Res. 2012. PMID: 22381566 Clinical Trial.
-
Accuracy and reproducibility of preoperative three-dimensional planning for total hip arthroplasty using biplanar low-dose radiographs : A pilot study.Orthop Traumatol Surg Res. 2017 Jun;103(4):531-536. doi: 10.1016/j.otsr.2017.03.001. Epub 2017 Mar 18. Orthop Traumatol Surg Res. 2017. PMID: 28323248
-
A prospective randomized controlled trial comparing CT-based planning with conventional total hip arthroplasty versus robotic arm-assisted total hip arthroplasty.Bone Joint J. 2024 Apr 1;106-B(4):324-335. doi: 10.1302/0301-620X.106B4.BJJ-2023-1045.R1. Bone Joint J. 2024. PMID: 38555946 Clinical Trial.
-
Femoral offset: anatomical concept, definition, assessment, implications for preoperative templating and hip arthroplasty.Orthop Traumatol Surg Res. 2009 May;95(3):210-9. doi: 10.1016/j.otsr.2009.03.010. Epub 2009 May 6. Orthop Traumatol Surg Res. 2009. PMID: 19423418 Review.
-
How to perform correct templating in total hip replacement.Musculoskelet Surg. 2023 Mar;107(1):19-28. doi: 10.1007/s12306-023-00772-3. Epub 2023 Jan 11. Musculoskelet Surg. 2023. PMID: 36630067 Review.
Cited by
-
Clinical application of preoperative volumetric measurement for interventional precise embolization in the treatment of hypersplenism caused by liver cirrhosis.BMC Surg. 2025 Jul 3;25(1):270. doi: 10.1186/s12893-025-03028-y. BMC Surg. 2025. PMID: 40611152 Free PMC article.
-
Comparison of traditional and robotic size prediction methods and stem positioning outcomes following prediction discrepancies in hip arthroplasty.J Robot Surg. 2025 Jul 21;19(1):409. doi: 10.1007/s11701-025-02594-w. J Robot Surg. 2025. PMID: 40690133
-
Two-Dimensional Preoperative Digital Templating is Less Accurate When Using a Collared Triple Taper Stem Versus a Single Taper Design.Arthroplast Today. 2025 Mar 11;32:101658. doi: 10.1016/j.artd.2025.101658. eCollection 2025 Apr. Arthroplast Today. 2025. PMID: 40130235 Free PMC article.
-
Novel Surgical Technique for Total Knee Arthroplasty Integrating Kinematic Alignment and Real-Time Elongation of the Ligaments Using the NextAR System.J Pers Med. 2024 Jul 26;14(8):794. doi: 10.3390/jpm14080794. J Pers Med. 2024. PMID: 39201986 Free PMC article.
-
[Study on assessment methods for acetabular cup size in total hip arthroplasty].Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2025 Feb 15;39(2):163-167. doi: 10.7507/1002-1892.202411005. Zhongguo Xiu Fu Chong Jian Wai Ke Za Zhi. 2025. PMID: 39971361 Free PMC article. Chinese.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Medical