. 2025 May;26(3):1076-1078.
doi: 10.1177/11297298241254255. Epub 2024 Jun 1.

Letter to the Editor re: "Systematic review and meta-analysis comparing Manta device and Perclose device for closure of large bore arterial access." J Vasc Access. 2024 Jan 8:11297298231222314

Affiliations

Letter to the Editor re: "Systematic review and meta-analysis comparing Manta device and Perclose device for closure of large bore arterial access." J Vasc Access. 2024 Jan 8:11297298231222314

Angela Di Giorgio et al. J Vasc Access. 2025 May.
No abstract available

PubMed Disclaimer

Conflict of interest statement

Declaration of conflicting interestsThe author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Figures

Figure 1.
Figure 1.
Ultrasound evaluation of large bore closure devices’ vascular complications. Common femoral artery (CFA)’s transverse (a) and longitudinal (b) ultrasound (B-mode) images showing MANTA’s normal ultrasound characteristics (a double-layered bright structure in the lumen formed by the poly-lactate toggle, and a perivascular hypoechogenic area corresponding to the collagen placement). Longitudinal ultrasound image (B-mode) (c) showing MANTA displacement leading to partial obstructive CFA thrombosis and hemodynamic stenosis on color Doppler ultrasound (d). Transverse (e) and longitudinal (f) ultrasound (B-mode) images showing no significant abnormalities of the CFA vessel wall in correspondence with Perclose, except for a hypoechoic shadow in the surrounding tissue, compatible with the introducer track. In presence of vessel diffuse atherosclerosis and vascular calcifications, the use of Perclose has resulted in a CFA dissection, with a visible intimal flap (highlighted in the image by cursors) (g) and acceleration of blood flow at this level on the color Doppler ultrasound image (h).

References

    1. Cheema T, Venero C, Jr, Champaneria S, et al.. Systematic review and meta-analysis comparing Manta device and Perclose device for closure of large bore arterial access. J Vasc Access. Epub ahead of print 8 January 2024. DOI: 10.1177/11297298231222314. - DOI - PubMed
    1. Schulz-Schüpke S, Helde S, Gewalt S, et al. Instrumental Sealing of Arterial Puncture Site- CLOSURE Device vs Manual Compression (ISAR-CLOSURE) Trial Investigators. Comparison of vascular closure devices vs manual compression after femoral artery puncture: the ISAR-CLOSURE randomized clinical trial. JAMA 2014; 312(19): 1981–1987. - PubMed
    1. Malkawi AH, Hinchliffe RJ, Holt PJ, et al.. Percutaneous access for endovascular aneurysm repair: a systematic review. Eur J Vasc Endovasc Surg 2010; 39(6): 676–682. - PubMed
    1. Van Mieghem NM, Latib A, van der Heyden J, et al.. Percutaneous plug-based arteriotomy closure device for large-bore access: a multicenter prospective study. JACC Cardiovasc Interv 2017; 10(6): 613–619. - PubMed
    1. d’Entremont MA, Alrashidi S, Seto AH, et al.. Ultrasound guidance for transfemoral access in coronary procedures: an individual participant-level data meta-analysis from the femoral ultrasound trialist collaboration. EuroIntervention 2024; 20(1): 66–74. - PMC - PubMed

LinkOut - more resources