E-vita OPEN NEO in the treatment of acute or chronic aortic pathologies: first interim results of the NEOS study
- PMID: 38830042
- PMCID: PMC11162351
- DOI: 10.1093/ejcts/ezae206
E-vita OPEN NEO in the treatment of acute or chronic aortic pathologies: first interim results of the NEOS study
Abstract
Objectives: The aim of this multicentre study was to demonstrate the safety and clinical performance of E-vita OPEN NEO Stent Graft System (Artivion, Inc.) in the treatment of aneurysm or dissection, both acute and chronic, in the ascending aorta, aortic arch and descending thoracic aorta.
Methods: In this observational study of 12 centres performed in Europe and in Asia patients were enrolled between December 2020 and March 2022. All patients underwent frozen elephant trunk using E-vita OPEN NEO Stent Graft System. Primary end point was the rate of all-cause mortality at 30 days and secondary end points included further clinical and safety data are reported up to 3-6 months postoperatively.
Results: A total of 100 patients (66.7% male; mean age, 57.7 years) were enrolled at 12 sites. A total of 99 patients underwent surgery using the E-vita OPEN NEO for acute or subacute type A aortic dissection (n = 37), chronic type A aortic dissection (n = 33) or thoracic aortic aneurysm (n = 29), while 1 patient did not undergo surgery. Device technical success at 24 h was achieved in 97.0%. At discharge, new disabling stroke occurred in 4.4%, while new paraplegia and new paraparesis was reported in 2.2% and 2.2%, respectively. Renal failure requiring permanent (>90 days) dialysis or hemofiltration at discharge was observed in 3.3% of patients. Between discharge and the 3-6 months visit, no patients experienced new disabling stroke, new paraplegia or new paraparesis. The 30-day mortality was 5.1% and the estimated 6-month survival rate was 91.6% (standard deviation: 2.9).
Conclusions: Total arch replacement with the E-vita OPEN NEO can be performed with excellent results in both the acute and chronic setting. This indicates that E-vita OPEN NEO can be used safely, including in the setting of acute type A aortic dissection.
Keywords: Antegrade cerebral perfusion; Aortic arch aneurysm; Aortic dissection; Distal aortic perfusion; E-vita OPEN NEO; Frozen elephant trunk procedure.
© The Author(s) 2024. Published by Oxford University Press on behalf of the European Association for Cardio-Thoracic Surgery.
Figures
References
-
- Czerny M, Schmidli J, Adler S, van den Berg JC, Bertoglio L, Carrel T. et al. Editor's Choice—Current Options and Recommendations for the Treatment of Thoracic Aortic Pathologies Involving the Aortic Arch: an Expert Consensus Document of the European Association for Cardio-Thoracic Surgery (EACTS) & the European Society for Vascular Surgery (ESVS). Eur J Vasc Endovasc Surg 2019;57:165–98. - PubMed
-
- Erbel R, Aboyans V, Boileau C, Bossone E, Bartolomeo RD, Eggebrecht H. et al. 2014 ESC Guidelines on the diagnosis and treatment of aortic diseases: document covering acute and chronic aortic diseases of the thoracic and abdominal aorta of the adult. The Task Force for the Diagnosis and Treatment of Aortic Diseases of the European Society of Cardiology (ESC). Eur Heart J 2014;35:2873–926. - PubMed
-
- Kato M, Ohnishi K, Kaneko M, Ueda T, Kishi D, Mizushima T. et al. New graft-implanting method for thoracic aortic aneurysm or dissection with a stented graft. Circulation 1996;94:II188–93. - PubMed
-
- Berger T, Weiss G, Voetsch A, Arnold Z, Kreibich M, Rylski B. et al. Multicentre experience with two frozen elephant trunk prostheses in the treatment of acute aortic dissection. Eur J Cardiothorac Surg 2019;56:572–8. - PubMed
-
- Mousavizadeh M, Bashir M, Jubouri M, Tan SZ, Borzeshi EZ, Ilkhani S. et al. Zone proximalization in frozen elephant trunk: what is the optimal zone for open intervention? A systematic review and meta-analysis. J Cardiovasc Surg (Torino) 2022;63:265–74. - PubMed
Publication types
MeSH terms
Grants and funding
LinkOut - more resources
Full Text Sources
Medical