. 2024 Dec;37(6):744-750.
doi: 10.1177/19714009241260802. Epub 2024 Jun 11.

The sensitivity of MIPs of 3D contrast-enhanced VIBE T1-weighted imaging for the detection of small brain metastases (≤ 5 mm) on 1.5 tesla MRI

Affiliations

The sensitivity of MIPs of 3D contrast-enhanced VIBE T1-weighted imaging for the detection of small brain metastases (≤ 5 mm) on 1.5 tesla MRI

Marco Parillo et al. Neuroradiol J. 2024 Dec.

Abstract

Objectives: To evaluate whether the use of Maximum Intensity Projection (MIP) images derived from contrast-enhanced 3D-T1-weighted volumetric interpolated breath-hold examination (VIBE) would allow more sensitive detection of small (≤5 mm) brain metastases (BM) compared with source as well as 2D-T1-weighted spin-echo (SE) images.

Methods: We performed a single center retrospective study on subjects with BM who underwent 1.5 tesla brain magnetic resonance imaging. Two readers counted the number of small BM for each of the seven sets of contrast-enhanced images created: axial 2D-T1-weighted SE, 3D-T1-weighted VIBE, 2.5 mm-thick-MIP T1-weighted VIBE, and 5 mm-thick-MIP T1-weighted VIBE; sagittal 3D-T1-weighted VIBE, 2.5 mm-thick-MIP T1-weighted VIBE, and 5 mm-thick-MIP T1-weighted VIBE. Total number of lesions detected on each image type was compared. Sensitivity, the average rates of false negatives and false positives, and the mean discrepancy were evaluated.

Results: A total of 403 small BM were identified in 49 patients. Significant differences were found: in the number of true positives and false negatives between the axial 2D-T1-weighted SE sequence and all other imaging techniques; in the number of false positives between the axial 2D-T1-weighted SE and the axial 3D-T1-weighted VIBE sequences. The two image types that combined offered the highest sensitivity were 2D-T1-weighted SE and axial 2.5 mm-thick-MIP T1-weighted VIBE. The axial 2D-T1-weighted SE sequence differed significantly in sensitivity from all other sequences.

Conclusion: MIP images did not show a significant difference in sensitivity for the detection of small BM compared with native images.

Keywords: Brain; central nervous system neoplasms; magnetic resonance imaging; maximum intensity projection; neoplasm metastasis; radiology; volumetric interpolated breath-hold examination.

PubMed Disclaimer

Conflict of interest statement

Declaration of conflicting interestsThe author(s) declared no potential conflicts of interest with respect to the research, authorship, and/or publication of this article.

Similar articles

Cited by

References

    1. Parillo M, Mallio CA, Pileri M, et al. Interrater reliability of Brain Tumor Reporting and Data System (BT-RADS) in the follow up of adult primary brain tumors: a single institution experience in Italy. Quant Imaging Med Surg 2023; 13(11): 7423–7431. - PMC - PubMed
    1. Aizer AA, Lamba N, Ahluwalia MS, et al. Brain metastases: a Society for Neuro-Oncology (SNO) consensus review on current management and future directions. Neuro Oncol 2022; 24(10): 1613–1646. - PMC - PubMed
    1. Mallio CA, Vaccarino F, Parillo M, et al. Artery of Huebner stroke chameleon presenting as central nervous system lymphoma: a case description. Quant Imaging Med Surg 2023; 13(11): 7646–7649. - PMC - PubMed
    1. Kaufmann TJ, Smits M, Boxerman J, et al. Consensus recommendations for a standardized brain tumor imaging protocol for clinical trials in brain metastases. Neuro Oncol 2020; 22(6): 757–772. - PMC - PubMed
    1. Danieli L, Riccitelli GC, Distefano D, et al. Brain tumor-enhancement visualization and morphometric assessment: a comparison of MPRAGE, SPACE, and VIBE MRI techniques. AJNR Am J Neuroradiol 2019; 40(7): 1140–1148. - PMC - PubMed

LinkOut - more resources