Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
. 2024 Sep;144(9):4125-4132.
doi: 10.1007/s00402-024-05336-z. Epub 2024 Jun 14.

Total knee replacement with an accelerometer-based, hand-held navigation system improves knee alignment: reliable in all patients

Affiliations

Total knee replacement with an accelerometer-based, hand-held navigation system improves knee alignment: reliable in all patients

Diego Alarcon Perico et al. Arch Orthop Trauma Surg. 2024 Sep.

Abstract

Background: Achieving adequate alignment has traditionally been an important goal in total knee arthroplasty to achieve long-term implant survival. While accelerometer-based hand-held navigation systems (ABN) has been introduced as a way to achieve alignment, there is a limited body of evidence on its accuracy, especially in patients under 65 years with differing etiologies for knee arthritis. This study aimed to assess the precision of a specific ABN system in restoring the mechanical axis and report surgical variables and complications, with particular attention to younger patients.

Methods: We conducted a retrospective review of 310 primary TKA performed with ABN from May 2016 to February 2021. The mean patient age was 67.4 (SD 8.9) years, with 43% under 65 years and mean body mass index of 33.2 (SD 6.8). The average surgical time was 96.8 min (57-171) and the average follow-up was 3.3 years (1.9-6.7). Data regarding length of stay, pain, range of motion (ROM), complications, and reinterventions were collected from the institutional joint arthroplasty registry and the medical records. Preoperative mechanical axis measurements and postoperative radiological data, including mechanical axis, component alignment and mechanical alignment outliers were analyzed.

Results: The mean preoperative mechanical axis was 175.4° (SD 7.6), with 248 knees (80%) in preoperative varus. The mean postoperative mechanical axis was 179.5° (SD 1.96) with 98% of knees falling within ± 3° of the neutral mechanical axis. Only 6 knees (2 varus, 4 valgus) fell outside the ± 3° range. And 3 knees (1 varus, 2 valgus) fell outside the ± 5° range. In the sagittal plane, 296 knees (95.5%) knees were within ± 3° of goal of 3 degrees of femoral flexion and 302 (97.4%) knees were within ± 2° of goal 1° of slope for tibial component. Far outliers (alignment outside ± 5° of targeted position) were found in 3 knees. Factors such as posttraumatic arthrosis, previous surgery, presence of retained hardware, and age below 65 years were not associated with increase in alignment outliers and far outliers. No complications related to the navigation system were observed. There were 22 complications and 20 reoperations, including 2 revisions for periprosthetic joint infection and 1 revision for flexion instability. Patients that required knee manipulation achieved an ultimate flexion of 110° (SD 14.1).

Conclusions: The ABN system proved to be user-friendly and accurate in reducing alignment outliers in both coronal and sagittal planes in all patient populations. It offers a straightforward navigation solution while preserving surgeon autonomy and the use of traditional surgical tools. These findings advocate for the integration of this navigation system as a valuable tool to enhance the precision of TKA surgery in all patient groups.

Keywords: Accelerometer based navigation; Complications; Knee alignment; OrthAlign; Total knee arthroplasty.

PubMed Disclaimer

Similar articles

References

    1. Kim YH et al (2014) The relationship between the survival of total knee arthroplasty and postoperative coronal, sagittal and rotational alignment of knee prosthesis. Int Orthop 38(2):379–385 - PubMed
    1. Liu HX et al (2016) Shorter survival rate in varus-aligned knees after total knee arthroplasty. Knee Surg Sports Traumatol Arthrosc 24(8):2663–2671 - PubMed
    1. Luyckx T et al (2023) Valgus alignment of the femoral component is associated with higher revision rates 10 years after TKA. Knee Surg Sports Traumatol Arthrosc 31(10):4171–4178 - PubMed
    1. Blakeney W et al (2019) Kinematic alignment in total knee arthroplasty better reproduces normal gait than mechanical alignment. Knee Surg Sports Traumatol Arthrosc 27(5):1410–1417 - PubMed
    1. An VVG et al (2019) Kinematic alignment is bone and soft tissue preserving compared to mechanical alignment in total knee arthroplasty. Knee 26(2):466–476 - PubMed

LinkOut - more resources