Accuracy of an autonomous dental implant robotic system in partial edentulism: A pilot clinical study
- PMID: 38890168
- DOI: 10.1007/s00784-024-05780-6
Accuracy of an autonomous dental implant robotic system in partial edentulism: A pilot clinical study
Abstract
Objectives: Robots are increasingly being used for surgical procedures in various specialties. However, information about the accuracy of robot-assisted dental implant surgery is lacking. This pilot clinical study aimed to investigate the accuracy of an autonomous dental implant robotic (ADIR) system in partially edentulous cases.
Material and methods: The ADIR system was used to place a total of 20 implants in 13 participants. Implant deviation from the planned positions was assessed to determine accuracy. The entry, apex, and angular deviations were described as means ± standard deviation. A two-sample t test was used to compare implant deviation between the flap and flapless groups and between maxillary and mandibular implants (α = .05).
Results: The entry, apex, and angular deviations were 0.65 ± 0.32 mm, 0.66 ± 0.34 mm, and 1.52 ± 1.01°, respectively, with no statistically significant difference between the flap and flapless approaches (P > .05). No adverse events were encountered in any of the participants.
Conclusions: DIR accuracy in this clinical series was comparable to that reported for static and dynamic computer-assisted implant surgery. Robotic computer-assisted implant surgery may be useful for dental implant placement, potentially improving the quality and safety of the procedure.
Clinical relevance: The findings of this study showed that the ADIR system could be useful for dental implant surgery.
Keywords: Accuracy; Autonomous dental implant robotic; Deviation; Implant.
© 2024. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature.
Similar articles
-
Accuracy of an autonomous dental implant robotic system in dental implant surgery.J Prosthet Dent. 2025 Mar;133(3):764-770. doi: 10.1016/j.prosdent.2024.07.020. Epub 2024 Aug 13. J Prosthet Dent. 2025. PMID: 39142994
-
Accuracy of an autonomous dental implant robotic system versus static guide-assisted implant surgery: A retrospective clinical study.J Prosthet Dent. 2025 Mar;133(3):771-779. doi: 10.1016/j.prosdent.2023.04.027. Epub 2023 Jun 7. J Prosthet Dent. 2025. PMID: 37291043
-
Accuracy of flapless surgery using an autonomous robotic system in full-arch immediate implant restoration: A case series.J Dent. 2024 Jun;145:105017. doi: 10.1016/j.jdent.2024.105017. Epub 2024 Apr 22. J Dent. 2024. PMID: 38657725
-
Accuracy of implant placement with computer-aided static, dynamic, and robot-assisted surgery: a systematic review and meta-analysis of clinical trials.BMC Oral Health. 2024 Mar 21;24(1):359. doi: 10.1186/s12903-024-04033-y. BMC Oral Health. 2024. PMID: 38509530 Free PMC article.
-
Accuracy of Dental Implant Placement using Robotics. A Systematic Review of in vitro Studies.Eur J Prosthodont Restor Dent. 2024 Sep 1;32(3):287-300. doi: 10.1922/EJPRD_2669Ravipati14. Eur J Prosthodont Restor Dent. 2024. PMID: 38691616
Cited by
-
Accuracy of robotic computer-assisted implant surgery in clinical studies: a systematic review and meta-analysis.BMC Oral Health. 2025 Apr 11;25(1):540. doi: 10.1186/s12903-025-05837-2. BMC Oral Health. 2025. PMID: 40217233 Free PMC article.
-
Accuracy and operation procedure of robotic computer-aided implant surgery.Sci Rep. 2025 Jan 2;15(1):307. doi: 10.1038/s41598-024-83997-x. Sci Rep. 2025. PMID: 39748004 Free PMC article.
-
Current Status and Future Perspectives of Robot-Assisted Dental Implant Surgery.Int Dent J. 2025 Jun;75(3):1608-1620. doi: 10.1016/j.identj.2025.02.020. Epub 2025 Mar 25. Int Dent J. 2025. PMID: 40138995 Free PMC article. Review.
References
-
- Buser D, Wittneben J, Bornstein MM, Grütter L, Chappuis V, Belser UC (2011) Stability of contour augmentation and esthetic outcomes of implant-supported single crowns in the esthetic zone: 3-year results of a prospective study with early implant placement postextraction. J Periodontol 82:342–349 - DOI - PubMed
-
- Buser D, Bornstein MM, Weber HP, Grütter L, Schmid B, Belser UC (2008) Early implant placement with simultaneous guided bone regeneration following single-tooth extraction in the esthetic zone: a cross-sectional, retrospective study in 45 subjects with a 2- to 4-year follow-up. J Periodontol 79:1773–1781 - DOI - PubMed
-
- D’Haese J, Ackhurst J, Wismeijer D, De Bruyn H (2000) Tahmaseb A (2017) Current state of the art of computer-guided implant surgery. Periodontol 73:121–133 - DOI
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous