Accuracy of an autonomous dental implant robotic system in partial edentulism: A pilot clinical study
- PMID: 38890168
- DOI: 10.1007/s00784-024-05780-6
Accuracy of an autonomous dental implant robotic system in partial edentulism: A pilot clinical study
Abstract
Objectives: Robots are increasingly being used for surgical procedures in various specialties. However, information about the accuracy of robot-assisted dental implant surgery is lacking. This pilot clinical study aimed to investigate the accuracy of an autonomous dental implant robotic (ADIR) system in partially edentulous cases.
Material and methods: The ADIR system was used to place a total of 20 implants in 13 participants. Implant deviation from the planned positions was assessed to determine accuracy. The entry, apex, and angular deviations were described as means ± standard deviation. A two-sample t test was used to compare implant deviation between the flap and flapless groups and between maxillary and mandibular implants (α = .05).
Results: The entry, apex, and angular deviations were 0.65 ± 0.32 mm, 0.66 ± 0.34 mm, and 1.52 ± 1.01°, respectively, with no statistically significant difference between the flap and flapless approaches (P > .05). No adverse events were encountered in any of the participants.
Conclusions: DIR accuracy in this clinical series was comparable to that reported for static and dynamic computer-assisted implant surgery. Robotic computer-assisted implant surgery may be useful for dental implant placement, potentially improving the quality and safety of the procedure.
Clinical relevance: The findings of this study showed that the ADIR system could be useful for dental implant surgery.
Keywords: Accuracy; Autonomous dental implant robotic; Deviation; Implant.
© 2024. The Author(s), under exclusive licence to Springer-Verlag GmbH Germany, part of Springer Nature.
References
-
- Buser D, Wittneben J, Bornstein MM, Grütter L, Chappuis V, Belser UC (2011) Stability of contour augmentation and esthetic outcomes of implant-supported single crowns in the esthetic zone: 3-year results of a prospective study with early implant placement postextraction. J Periodontol 82:342–349 - DOI - PubMed
-
- Buser D, Bornstein MM, Weber HP, Grütter L, Schmid B, Belser UC (2008) Early implant placement with simultaneous guided bone regeneration following single-tooth extraction in the esthetic zone: a cross-sectional, retrospective study in 45 subjects with a 2- to 4-year follow-up. J Periodontol 79:1773–1781 - DOI - PubMed
-
- D’Haese J, Ackhurst J, Wismeijer D, De Bruyn H (2000) Tahmaseb A (2017) Current state of the art of computer-guided implant surgery. Periodontol 73:121–133 - DOI
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Miscellaneous