Direct enzyme immunoassay for detection of specific IgG antibody to rubella virus by use of labelled antigen
- PMID: 3894404
- DOI: 10.1016/0166-0934(85)90039-4
Direct enzyme immunoassay for detection of specific IgG antibody to rubella virus by use of labelled antigen
Abstract
A new solid phase enzyme immunoassay (EIA) for detection of rubella-specific immunoglobulin G (IgG) antibody was developed. The test uses polystyrene microtiter strips coated with rabbit anti-human IgG immunoglobulins as the solid phase and an enzyme-labelled semipurified rubella antigen as indicator. The direct EIA was compared with hemagglutination inhibition (HI), single radial hemolysis (SRH), radioimmunoassay (RIA) and time-resolved fluoroimmunoassay (TR-FIA) using 52 serum specimens from patients with remote rubella infection. The overall agreement of direct EIA with HI was 96.1%, with SRH and RIA 98.1% and with TR-FIA 100%. The linear regression coefficient varied from 0.77 to 0.91, the best being obtained with direct EIA and SRH. The direct EIA was also suitable for diagnosis of acute infections, as a significant increase in antibody levels was detected in all paired specimens tested from patients with acute rubella infection. The sensitivity and were comparable to those of the assays employed. An advantage of the present assay is that the same method and same labelled antigen can be used to test for different classes of antibody using simply a solid phase with capture antibodies of different chain specificity.
Similar articles
-
Time-resolved fluoroimmunoassay: a new test for rubella antibodies.J Clin Microbiol. 1982 Nov;16(5):920-5. doi: 10.1128/jcm.16.5.920-925.1982. J Clin Microbiol. 1982. PMID: 6759532 Free PMC article.
-
Detection of specific immunoglobulin M antibody to rubella virus by use of an enzyme-labeled antigen.J Clin Microbiol. 1985 Jun;21(6):963-8. doi: 10.1128/jcm.21.6.963-968.1985. J Clin Microbiol. 1985. PMID: 3891775 Free PMC article.
-
Hemagglutination inhibition, single radial hemolysis, and ELISA tests for the detection of IgG and IgM to rubella virus.J Med Virol. 1980;5(4):273-86. doi: 10.1002/1096-9071(1980)5:4<273::aid-jmv1890050403>3.0.co;2-j. J Med Virol. 1980. PMID: 7014780
-
[Diagnostic tests: Rubella virus].Nihon Rinsho. 2005 Jul;63 Suppl 7:356-8. Nihon Rinsho. 2005. PMID: 16111273 Review. Japanese. No abstract available.
-
Measurement of cytokine antibodies. Test development.Biotherapy. 1997;10(1):87-92. doi: 10.1007/BF02678219. Biotherapy. 1997. PMID: 9261552 Review.