Granulomatous reaction and cystic bony destruction associated with high wear rate in a total knee prosthesis
- PMID: 3896606
Granulomatous reaction and cystic bony destruction associated with high wear rate in a total knee prosthesis
Abstract
This is a case report of a Cloutier total knee arthroplasty that was removed from an obese, large, 66-year-old man three years after implantation because of aseptic loosening. At surgery there was a thickened black synovium and black cystic areas in the exposed bone. The articulating surfaces of the tibial component, which is made from Poly Two (a carbon polyethylene composite), were grossly abraded, and the supporting metal tray was broken in two. The femoral component showed signs of abnormal wear at the places where it was articulating with the displaced tibial component. The tissues showed a granulomatous reaction with marrow fibrosis and cystic destruction of bone. It is postulated that aseptic loosening was accelerated by a granulomatous response to overload, abrasion, and local dissemination of particles.
Similar articles
-
Failure of a knee prosthesis accelerated by shedding of beads from the porous metal surface.Clin Orthop Relat Res. 1990 Sep;(258):204-8. Clin Orthop Relat Res. 1990. PMID: 2203570
-
[Case report of extensive metallosis in extra-articular tissues after unicompartmental knee joint replacement].Acta Chir Orthop Traumatol Cech. 2003;70(1):47-50. Acta Chir Orthop Traumatol Cech. 2003. PMID: 12764951 Czech.
-
Granulomatous reaction and cystic bony destruction associated with high wear rate in a total knee prosthesis.Clin Orthop Relat Res. 1987 Mar;(216):309. Clin Orthop Relat Res. 1987. PMID: 3815965 No abstract available.
-
Popliteal cyst in a patient with total knee arthroplasty: a case report and review of the literature.J Med Liban. 2001 Nov-Dec;49(6):347-50. J Med Liban. 2001. PMID: 12744639 Review.
-
[Wear particles: key to aseptic prosthetic loosening?].Pathologe. 2006 Nov;27(6):447-60. doi: 10.1007/s00292-006-0868-4. Pathologe. 2006. PMID: 17024444 Review. German.
Cited by
-
Ultra-High-Molecular-Weight Polyethylene in Hip and Knee Arthroplasties.Materials (Basel). 2023 Mar 7;16(6):2140. doi: 10.3390/ma16062140. Materials (Basel). 2023. PMID: 36984020 Free PMC article. Review.
-
Analysis of the effect of pressure on compression moulding of UHMWPE.J Mater Sci Mater Med. 1998 Mar;9(3):165-72. doi: 10.1023/a:1008871720389. J Mater Sci Mater Med. 1998. PMID: 15348906
-
[What effect does of obesity have on the outcome of total hip and knee arthroplasty. Review of the literature].Orthopade. 2005 Jul;34(7):664-7. doi: 10.1007/s00132-005-0815-8. Orthopade. 2005. PMID: 15931520 Review. German.
-
Tibial tray fracture in a modern prosthesis with retrieval analysis.Arthroplast Today. 2018 Feb 10;4(2):143-147. doi: 10.1016/j.artd.2017.12.005. eCollection 2018 Jun. Arthroplast Today. 2018. PMID: 29896542 Free PMC article.
-
Fracture of the tibial tray following PCA knee replacement. A report of two cases.Arch Orthop Trauma Surg. 1994;113(6):330-3. doi: 10.1007/BF00426182. Arch Orthop Trauma Surg. 1994. PMID: 7833211
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources